The stability of three different cementless tibial components: A randomized radiostereometric study in 45 knee arthroplasty patients
- 1 January 1995
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 66 (1) , 21-27
- https://doi.org/10.3109/17453679508994633
Abstract
We conducted a prospective, randomized study of 45 patients to evaluate 3 different uncemented tibial component designs in total knee arthroplasty. The stability of the components was assessed by radiostereometry (RSA), both as migration during 2 years and as inducible displacement at 2 years. The PCA resurfacing, the Tricon stem and the Tricon-M prosthesis groups showed a similar level of migration at 2 years, about 1.4 mm. In response to externally applied rotatory forces, the Tricon groups rotated more than the PCA group, interpreted as a consequence of the more conforming articular surface in the Tricon design. The series was divided into one group of continuously migrating prostheses with a poor prognosis (unstable, one third) and another group of prostheses in which migration stopped after 1 year (stable, two thirds). With this classification, no differences between the prostheses design groups were revealed. However, the unstable group showed a larger inducible displacement by provocation, an association hitherto not established.Keywords
This publication has 12 references indexed in Scilit:
- Early migration and late aseptic failure of proximal femoral prosthesesThe Journal of Bone and Joint Surgery. British volume, 1994
- Failed innovation in total hip replacement: Diagnosis and proposals for a cureActa Orthopaedica, 1993
- Early postoperative fixation of tibial components: An in vivo roentgen stereophotogrammetric analysisJournal of Orthopaedic Research, 1993
- Polyethylene wear of the PCA unicompartmental kneeActa Orthopaedica, 1992
- Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prosthesesThe Journal of Bone and Joint Surgery. British volume, 1992
- A comparative study of uncemented tibial componentsThe Journal of Arthroplasty, 1990
- Chronic lateral instability of the ankle Roentgen stereophotogrammetry of talar positionActa Orthopaedica, 1989
- Inherent laxity in total knee prosthesesThe Journal of Arthroplasty, 1987
- Observations on the Effect of Movement on Bone Ingrowth into Porous-Surfaced ImplantsClinical Orthopaedics and Related Research, 1986
- The fixation of a proximal tibial polyethylene prosthesis without cementThe Journal of Bone and Joint Surgery. British volume, 1982