Clinical Utility of Aortic Pulses and Pressures Calculated From Applanated Radial-Artery Pulses
- 1 August 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Hypertension
- Vol. 42 (2) , 150-155
- https://doi.org/10.1161/01.hyp.0000084051.34269.a9
Abstract
Brachial artery cuff blood pressures are but approximations of central aortic pressures. The actual pressures against which the left ventricle must pump would be useful clinical information if obtained noninvasively. Our aim was to determine the clinical utility of aortic pulses and pressures calculated from noninvasively obtained radial-artery pulses. Radial-arterial pulses were recorded by applanation and calibrated with arm/cuff oscillometric pressures. Aortic pulses and pressures were calculated from the radial pulses by Fourier analysis and transfer functions. These calculated aortic pulses were compared with directly recorded aortic pulses by a transducer-tip catheter in a series of 50 patients undergoing cardiac catheterization. The correlation coefficient ( r ) of the measured versus the calculated aortic systolic blood pressure was +0.89, but the scatter was large (standard deviation of the differences=±11.3 mm Hg). The pulse pressure correlations were less good ( r =+0.79) and also had a large scatter (±13.6 mm Hg). The average calculated pulse pressure was 11.5 mm Hg lower than the measured value because the cuff diastolic blood pressures, used to calibrate the radial pulses, were systematically higher than those in the aorta (8.9 mm Hg). Multivariable analysis incorporating height, age, heart rate, and ejection fraction as additional, independent variables eliminated mean differences between the new “predicted” and measured pressures, significantly improved correlation coefficients, and reduced the scatter. However, the improvements were small. The inaccuracy of the oscillometric cuff method for measuring arm blood pressure appears to be the limiting factor in the prediction of clinically useful, noninvasive aortic pressures.Keywords
This publication has 15 references indexed in Scilit:
- Pulse wave analysisBritish Journal of Clinical Pharmacology, 2001
- Validation of the Welch Allyn ‘Vital Signs’ oscillometric blood pressure monitorJournal of Human Hypertension, 2001
- Parametric model derivation of transfer function for noninvasive estimation of aortic pressure by radial tonometryIEEE Transactions on Biomedical Engineering, 1999
- Comparison of oscillometric and intra-arterial blood pressure and pulse measurementJournal of Medical Engineering & Technology, 1994
- An analysis of the relationship between central aortic and peripheral upper limb pressure waves in manEuropean Heart Journal, 1993
- Blood pressure measurement during transport A comparison of direct and oscillotonometric readings in critically ill patientsAnaesthesia, 1990
- The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systemsJournal Of Hypertension, 1990
- A Comparison of Two Automated Indirect Arterial Blood Pressure MetersAnesthesiology, 1985
- Influence of ventricular ejection on the relationship between central aortic and brachial pressure pulse in manCardiovascular Research, 1970
- Blood Pressure Measurement in ShockJAMA, 1967