Adaptive QRS Detection: A Study of Performance
- 1 June 1985
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. BME-32 (6) , 392-401
- https://doi.org/10.1109/tbme.1985.325465
Abstract
QRS detectors are often evaluated in terms of statistical measures, e.g., rate of false and true detections. These measures are usually calculated for a detector with fixed parameter values. In this paper, detector properties are studied by varying those parameters which seem to have the most profound effect on performance. By applying techniques known from optimal estimation to a stochastic model for the ECG, two different detectors have been derived. The detector structures are described in detail. In order to study and compare the performance of the two detectors, a database of "difficult" ECG's has been collected, containing various "artifacts" of physiological and electrical/mechanical origin. The performance of this database is presented in terms of receiver-operating characteristics (ROC's), and several examples are discussed.Keywords
This publication has 10 references indexed in Scilit:
- Optimal QRS detectorMedical & Biological Engineering & Computing, 1983
- Maximum likelihood detection and estimation of Bernoulli - Gaussian processesIEEE Transactions on Information Theory, 1982
- Adaptive QRS Detection Based on Maximum A Posteriori EstimationIEEE Transactions on Biomedical Engineering, 1982
- Estimation of pulse heights and arrival timesAutomatica, 1980
- Abdominal-Lead Fetal Electrocardiographic R-Wave Enhancement for Heart Rate DeterminationIEEE Transactions on Biomedical Engineering, 1980
- A Digital Filter for the QRS Complex DetectionIEEE Transactions on Biomedical Engineering, 1979
- Resolution of Overlapping Echo Signals in Sonar/Radar SystemsIEEE Transactions on Communications, 1979
- Microprocessor detection of electrocardiogram R-wavesJournal of Medical Engineering & Technology, 1979
- Automated cardiac dysrhythmia analysisProceedings of the IEEE, 1979
- Automatic real time arrhythmia monitoring in the intensive coronary care unitThe American Journal of Cardiology, 1977