Alveolar pressure monitoring: an evaluation in a lung model and in patients with acute lung injury
- 11 April 2003
- journal article
- research article
- Published by Springer Nature in Intensive Care Medicine
- Vol. 29 (6) , 955-962
- https://doi.org/10.1007/s00134-003-1730-2
Abstract
We evaluated an algorithm for continuous on-line monitoring of alveolar pressure over time in a lung model with lower and upper inflection points and variable resistance ratios and in patients with acute lung injury. The algorithm is based on "static" pressure/volume curves obtained from tracheal pressure measurements under dynamic conditions. Experimental and clinical evaluation of algorithm in a university hospital laboratory and intensive care unit. Ten patients undergoing postoperative respiratory therapy (feasibility of tracheal measurement) and ten patients with acute lung injury undergoing ventilator treatment (evaluation of algorithm). Direct tracheal pressure measurements with a catheter inserted through the endotracheal tube. Comparison of measured alveolar and the dynostatic alveolar pressure vs. time in a lung model with changes in five ventilatory parameters. Examples of clinical monitoring are reported. In the model there was excellent agreement between alveolar pressures obtained by the algorithm, the dynostatic alveolar pressure, and measured alveolar pressure at all ventilator settings. For inspiratory/expiratory resistance ratios between 1:2.1–2.1:1, the dynostatic alveolar pressure was within ±1.5 cm H2O of measured alveolar pressure. In patients the technique for direct tracheal pressure measurement using a catheter inserted through the endotracheal tube functioned satisfactorily with intermittent air flushes for cleansing. Using a thin tracheal pressure catheter inserted through the endotracheal tube alveolar pressure allows continuous bedside monitoring with ease and precision using the dynostatic algorithm. The method is unaffected by tube and connector geometry or by secretions.Keywords
This publication has 33 references indexed in Scilit:
- Ventilation with Lower Tidal Volumes as Compared with Traditional Tidal Volumes for Acute Lung Injury and the Acute Respiratory Distress SyndromeNew England Journal of Medicine, 2000
- Prospective, randomized, controlled clinical trial comparing traditional versus reduced tidal volume ventilation in acute respiratory distress syndrome patientsCritical Care Medicine, 1999
- Gradual Reduction of Endotracheal Tube Diameter during Mechanical Ventilation via Different Humidification DevicesAnesthesiology, 1996
- Respiratory mechanics in patients ventilated for critical lung diseaseEuropean Respiratory Journal, 1996
- Comparison of six methods to calculate airway resistance during mechanical ventilation in adultsJournal of Clinical Monitoring and Computing, 1993
- Maneuver-Free Determination of Compliance and Resistance in Ventilated ARDS PatientsChest, 1992
- Respiratory Resistance in Mechanically Ventilated Patients: The Effects of PEEPPublished by Springer Nature ,1992
- A method for studying the static volume-pressure curves of the respiratory system during mechanical ventilationJournal of Critical Care, 1989
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Total Respiratory Pressure-Volume Curves in the Adult Respiratory Distress SyndromeChest, 1984