Determination of Ribavirin in Serum Using Highly Selective Solid-Phase Extraction and High-Performance Liquid Chromatography

Abstract
A rapid assay for determination of ribavirin in serum using solid-phase extraction (SPE), high-performance liquid chromatography (HPLC), and UV-detection was developed. The SPE uses phenylboronic acid columns with an approximately 100% recovery for ribavirin. The concentration-peak area relation was linear (r > 0.995), from 1 to 64 microM in 100 microL serum. The limit of detection was 0.1 microM. The intraassay CV was 3.2% at treatment levels (9.7 microM) and 11.5% at 0.4 microM. The method is used to monitor patients undergoing ribavirin treatment for hepatitis C (HCV). Samples from HCV-infected patients with and without renal dysfunction have been analyzed without interference of endogenous compounds. It is concluded that the method is useful for routine therapeutic drug monitoring.

This publication has 11 references indexed in Scilit: