Experimental dosimetry of a32P catheter-based endovascular brachytherapy source
- 14 July 2003
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 48 (15) , 2283-2296
- https://doi.org/10.1088/0031-9155/48/15/303
Abstract
The experimental dosimetry in a water phantom of a 32P linear source, 20 mm in length, used for the brachytherapy of coronary vessels is reported. The source content activity, A, was determined by means of a calibrated well ion-chamber and the value was compared with the contained activity reported in the manufacturer's certification. In this field of brachytherapy dosimetry, radiochromic film supplies a high enough spatial resolution. A highly sensitive radiochromic film, that presents only one active layer, was used in this work for the source dosimetry in a water phantom. The radiochromic film was characterized by electron beams produced by a clinical linac. A Monte Carlo calculation of beta spectra in water at different distances along the source transverse bisector axis allowed to take into account the low dependence of film response from the electron beam energy. The adopted experimental set-up, with the source in its catheter positioned on the film plane inside the water phantom, supplies accurate dosimetric information. The measured dose rate to water per unit of source activity at reference distance, (r0, θ0)/A, in units of cGy s−1 GBq−1, was in agreement with the value reported in the manufacturer's certification within the experimental uncertainty. The radial dose function, g(r), is in good agreement with the literature data. The anisotropy function F(r, θ) is also reported. The analysis of the dose profile obtained at 2 mm from the source longitudinal axis shows that the uniformity is within 10% along 75% of the 20 mm treatment length. The adopted experimental set-up seems to be adequate for the quality control procedure of the dose homogeneity distribution in the water medium.Keywords
This publication has 12 references indexed in Scilit:
- A standard dosimetry procedure for192Ir sources used for endovascular brachytherapyPhysics in Medicine & Biology, 2002
- Dosimetry characterization of 32P catheter-based vascular brachytherapy source wireMedical Physics, 2000
- Verification of absolute dose rates for intravascular brachytherapy beta sourcesPhysics in Medicine & Biology, 2000
- Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 versionMedical Physics, 2000
- Monte Carlo calculations of dose distributions around 32P and 198Au stents for intravascular brachytherapyMedical Physics, 1999
- Intravascular brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 60Medical Physics, 1999
- Quantitative evaluation of radiochromic film response for two‐dimensional dosimetryMedical Physics, 1997
- Novel Radiochromic Films for Clinical DosimetryRadiation Protection Dosimetry, 1996
- Calibration and use of Plane-Parallel Ionization Chambers for the Determination of Absorbed Dose in Electron BeamsActa Radiologica: Oncology, 1981