A standard dosimetry procedure for192Ir sources used for endovascular brachytherapy
- 12 November 2002
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 47 (23) , 4205-4221
- https://doi.org/10.1088/0031-9155/47/23/307
Abstract
The experimental dosimetry of a high dose rate (HDR) 192Ir source used for the brachytherapy of peripheral vessels is reported. The direct determination of the reference air kerma rate r agrees, within the experimental uncertainty, with the results obtained by a well ionization chamber calibrated at the NIST and the manufacturer's certification. A highly sensitive (HS) radiochromic film (RCF), that presents only one active layer, was used for the source dosimetry in a water phantom. The adopted experimental set-up, with the source in its catheter positioned on the RCF plane, seems to have given better accuracy of the RCF optical density measurements. The agreement between the measurement of the dose rate constant DKr(10 mm, π/2) and the literature data confirmed the coherence of the HS RCF calibration obtained by the kerma in air measurements. The RCF measurements supplied dosimetric information about the dose to water per reference air kerma rate (r, θ)/r along the source transverse bisector axis, the radial dose function g(r) and the anisotropy function F(r, θ). The value (2 mm, π/2)/r = 22.4 ± 1.2 cGy h−1/(μGy h−1) is supplied with a dose uncertainty that is essentially due to the indeterminacy of the source position in the catheter. The data of the radial and anisotropy functions have been compared with Monte Carlo determinations reported in the literature.Keywords
This publication has 19 references indexed in Scilit:
- Measurements and Monte Carlo calculations to determine the absolute detector response of radiochromic film for brachytherapy dosimetryMedical Physics, 2001
- Radiochromic film dosimetry of a low energy proton beamMedical Physics, 2000
- Intravascular brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 60Medical Physics, 1999
- Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55Medical Physics, 1998
- Influence of source geometry and materials on the transverse axis dosimetry of brachytherapy sourcesPhysics in Medicine & Biology, 1998
- Dosimetric characteristics of a new high‐intensity source for remote afterloadingMedical Physics, 1997
- Quantitative evaluation of radiochromic film response for two‐dimensional dosimetryMedical Physics, 1997
- Novel Radiochromic Films for Clinical DosimetryRadiation Protection Dosimetry, 1996
- Dose for Nonelectronic Equilibrium ConditionsRadiation Research, 1958