Toward quantitative fluorescence measurements with multicolor flow cytometry
Open Access
- 28 December 2007
- journal article
- research article
- Published by Wiley in Cytometry Part A
- Vol. 73A (4) , 279-288
- https://doi.org/10.1002/cyto.a.20507
Abstract
A procedure is presented for calibrating the output of a multicolor flow cytometer in units of antibodies bound per cell (ABC). The procedure involves two steps. First, each of the fluorescence channels of the flow cytometer is calibrated using Ultra Rainbow beads with assigned values of equivalent number of reference fluorophores (ERF). The objective of this step is to establish a linear relation between the fluorescence signal in a given fluorescence channel of multicolor flow cytometers and the value of ERF. The second step involves a biological standard such as a lymphocyte with a known number of antibody binding sites (e.g., CD4 binding sites). The biological standard is incubated with antibodies labeled with one type of fluorophores for a particular fluorescence channel and serves to translate the ERF scale to an ABC scale. A significant part of the two‐step calibration procedure involves the assignment of ERF values to the different populations of Ultra Rainbow beads. The assignment of ERF values quantifies the relative amount of embedded fluorophore mixture in each bead population. It is crucial to insure that the fluorescence signal in a given range of fluorescence emission wavelengths is related linearly to the assigned values of ERF. The biological standard has to posses a known number of binding sites for a given antibody. In addition, this antibody has to be amenable to labeling with different types of fluorophores associated with various fluorescence channels. The present work suggests that all of the requirements for a successful calibration of a multicolor flow cytometer in terms of ABC values can be fulfilled. The calibration procedure is based on firm scientific foundations so that it is easy to envision future improvements in accuracy and ease of implementation. Published 2007 Wiley‐Liss, Inc.Keywords
This publication has 15 references indexed in Scilit:
- Comparative Analysis of Flow Cytometric Techniques in Assessment of ZAP-70 Expression in Relation toIgVHMutational Status in Chronic Lymphocytic LeukemiaAmerican Journal of Clinical Pathology, 2007
- Comparison of fluorescein and phycoerythrin conjugates for quantifying CD20 expression on normal and leukemic B‐cellsCytometry Part B: Clinical Cytometry, 2006
- Quality assurance for polychromatic flow cytometryNature Protocols, 2006
- Enumeration of peripheral lymphocyte subsets using 6 vs. 4 color staining: A clinical evaluation of a new flowcytometerCytometry Part B: Clinical Cytometry, 2005
- Critical Path Workshop on the Development of Cellular and Gene Therapy ProductsMolecular Therapy, 2005
- Quantitating fluorescence intensity from fluorophore: Assignment of MESF valuesJournal of Research of the National Institute of Standards and Technology, 2005
- ZAP-70 Expression as a Surrogate for Immunoglobulin-Variable-Region Mutations in Chronic Lymphocytic LeukemiaNew England Journal of Medicine, 2003
- Crystal Structure of Allophycocyanin from Red AlgaePorphyra yezoensis at 2.2-Å ResolutionJournal of Biological Chemistry, 1999
- Quantitation of CD38 activation antigen expression on CD8+ T cells in HIV-1 infection using CD4 expression on CD4+ T lymphocytes as a biological calibratorCytometry, 1998
- Isolation, crystallization, crystal structure analysis and refinement of allophycocyanin from the cyanobacterium Spirulina platensis at 2.3 Å resolutionJournal of Molecular Biology, 1995