Pulsation artifact in short TR MR imaging and angiography: Exacerbation with signal averaging
- 1 September 1994
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 4 (5) , 709-718
- https://doi.org/10.1002/jmri.1880040514
Abstract
Averaging the signals from more than one excitation per phase‐encoding view increases the signal‐to‐noise ratio and, in conventional spin‐echo magnetic resonance imaging, reduces most motion artifacts. To determine the effects of signal averaging on two‐dimensional gradient‐echo images, acquisitions with different TRs and with no averaging versus multiple‐signal averaging were compared in a pulsatile flow phantom and the human abdominal aorta. Intraview (each view repeated before changing the phase‐encoding value) and interview (obtaining all views sequentially and then repeating the entire set) averaging methods were used. Pulsation artifacts were present on all images of the flow phantom and the aorta. Intraview signal averaging, the method most commonly used, exacerbated rather than ameliorated pulsation artifacts with short TR sequences. Pulsation artifacts on two‐dimensional images obtained with a short TR can be minimized by completing the acquisition as rapidly as possible, avoiding signal averaging. If signal averaging is used for short TR images, it should be interview averaging.Keywords
This publication has 13 references indexed in Scilit:
- Pulsatile flow artifacts in two‐dimensional time‐of‐flight MR angiography: Initial studies in elastic models of human carotid arteriesJournal of Magnetic Resonance Imaging, 1993
- Aortic ghost artifact in ultrashort te multislice gradient echo MR images is not increased by paramagnetic enhancementMagnetic Resonance in Medicine, 1993
- Blood flow imaging through detection of temporal variations in magnetization.Radiology, 1992
- T1-weighted snapshot gradient-echo MR imaging of the abdomen.Radiology, 1991
- Ineffectiveness of averaging for reducing motion artifacts in half-fourier MR imagingJournal of Magnetic Resonance Imaging, 1991
- Flow‐independent magnetic resonance projection angiographyMagnetic Resonance in Medicine, 1991
- Optimization of Gradient-Echo Imaging Parameters for Intracaval Filters and Trapped ThromboemboliRadiology, 1990
- Nuclear magnetic resonance signal from flowing nuclei in rapid imaging using gradient echoesMedical Physics, 1988
- Motion artifact reduction with fast spin-echo imaging.Radiology, 1987
- Suppression of respiratory motion artifacts in magnetic resonance imagingMedical Physics, 1986