Profile distortions during secondary ion mass spectrometry analyses of resistive layers due to electron stimulated desorption and charging
- 1 October 1986
- journal article
- Published by AIP Publishing in Journal of Applied Physics
- Vol. 60 (7) , 2573-2579
- https://doi.org/10.1063/1.337123
Abstract
The problem of charge compensation during SIMS depth profiling of resistive-layer structures on insulating substrates is discussed with reference to BF2 and As-implanted silicon-on-sapphire (SOS). Accurate positive ion fluorine profiles (from the BF2 in SOS samples) could not be obtained using electron beam charge compensation, due to an intense electron stimulated desorption (ESD) signal at mass 19. This species is shown to be F+. Evaporation of a gold-surface-edge electrode, 50 nm thick, represents an alternative charge compensation technique, providing a conduction path from crater edge to sample holder that bypasses the insulating substrate. Fluorine profiles can thus be obtained to within a few hundred angstroms of the sapphire. Such an electrode may be of more general use as it simplifies the setting up of electron beam charge compensation, where profiles are to be taken through a conductor and into an insulator, facilitating correct charge compensation at both sides of the interface. Accurate arsenic profiles (from the As in SOS samples) also require careful charge compensation, particularly in quadrupole based machines. Small changes in surface potential can lead to the illusion of dopant pile up (at an interface, for example) as Si2O is swept into the bandpass of the instrument. It is demonstrated that the intercomparison of mass channels, tuned to the same mass but different starting potentials, is an invaluable aid in the detection of such artifacts. The comparison of matrix-channel-ion images obtained at different stages in the profile is also useful and shows that the charging of the crater base does not occur uniformly.This publication has 11 references indexed in Scilit:
- An electrostatic iris for use in quadrupole SIMS instrumentsNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1984
- Observations of electron stimulated desorption during SIMS profiling of insulatorsVacuum, 1984
- Quantification of dopant implants in oxidized silicon on sapphire using secondary-ion mass spectrometryJournal of Applied Physics, 1983
- Experimental study of electrode materials for use in a cold-cathode oxygen dischargeInternational Journal of Mass Spectrometry and Ion Physics, 1983
- Improved secondary-ion extraction in a quadrupole-based ion microprobeInternational Journal of Mass Spectrometry and Ion Physics, 1982
- Ion-stimulated desorption of positive halogen ionsPhysical Review B, 1981
- Primary-ion charge compensation in SIMS analysis of insulatorsJournal of Applied Physics, 1979
- On the mechanism of cluster emission in sputteringPhysics Letters A, 1979
- High-sensitivity depth profiling of arsenic and phosphorus in silicon by means of SIMSApplied Physics Letters, 1976
- The origin of spurious peaks in mass spectraJournal of Physics D: Applied Physics, 1976