Evaluation of oxygenators and centrifugal pumps for long-term pediatric extracorporeal membrane oxygenation
- 1 September 2007
- journal article
- research article
- Published by SAGE Publications in Perfusion
- Vol. 22 (5) , 323-326
- https://doi.org/10.1177/0267659107086270
Abstract
Objective : Two extracorporeal membrane oxygenation (ECMO) circuits for children under 10 kg were evaluated and compared for plasma leakage, hemolysis, blood transfusions, and durability. Methods: Group A (n = 20) was supported by ECMO circuits with the Minimax oxygenator and the Biomedicus centrifugal pump. Group B (n = 10) was supported by ECMO circuits with the Lilliput 2 ECMO oxygenator and the Rotaflow centrifugal pump. Results: ECMO circuit durability, as measured by oxygenator lifespan, was significantly better in Group B than in Group A (p = 0.04). There was significantly lower hemolysis, measured by plasma free hemoglobin, in Group B (p = 0.019), and patients in Group B had significantly less need for antithrombin III transfusion (p = 0.004). No plasma leakage was observed in Group B oxygenators, but plasma leakage was observed in all Group A oxygenators. Conclusion: The combination of a Rotaflow centrifugal pump and Lilliput 2 ECMO oxygenator in pediatric ECMO circuits improved durability and reduced circuit-induced hemolysis. This improvement may be due to the low priming volume, the oxygenator's plasma leakage resistance, the suspended rotor of the centrifugal pump, or a combination of these factors.Perfusion (2007) 22, 323—326.Keywords
This publication has 16 references indexed in Scilit:
- Evaluation of Plasma Resistant Hollow Fiber Membranes For Artificial LungsAsaio Journal, 2004
- Experience with the Jostra Rotaflow and QuadroxD oxygenator for ECMOPerfusion, 2004
- Extracorporeal membrane oxygenation using a centrifugal pump and a servo regulator to prevent negative inlet pressureThe Annals of Thoracic Surgery, 1997
- Hemolysis in Different Centrifugal PumpsArtificial Organs, 1997
- Seal‐less Centrifugal Blood Pump with Magnetically Suspended Rotor: Rot‐a‐FlotArtificial Organs, 1995
- Effect of Stationary Guiding Vanes on Improvement of the Washout Behind the Rotor in Centrifugal Blood PumpsAsaio Journal, 1992
- Laboratory Experience with a Novel, Non-occlusive, Pressure-regulated Peristaltic Blood PumpAsaio Journal, 1992
- Pump-induced haemolysis: is the constrained vortex pump better or worse than the roller pump?Perfusion, 1992
- Hemolysis during long-term extracorporeal membrane oxygenationThe Journal of Pediatrics, 1989
- Extracorporeal Circulation for Support of Palliative Cardiac Surgery in InfantsThe Annals of Thoracic Surgery, 1970