Hydrophilic coating aids radial sheath withdrawal and reduces patient discomfort following transradial coronary intervention: A randomized double‐blind comparison of coated and uncoated sheaths
- 20 May 2003
- journal article
- clinical trial
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 59 (2) , 161-164
- https://doi.org/10.1002/ccd.10444
Abstract
Radial artery spasm may cause severe discomfort during radial artery sheath removal. A hydrophilic‐coated sheath may reduce the force required to remove a radial sheath. This force may be quantified using an automatic pullback device (APD). The objective of this study was to assess if a hydrophilic coating reduces the required force and discomfort associated with removal of a radial sheath following transradial coronary intervention. Ninety patients undergoing percutaneous coronary intervention via the radial artery were randomly assigned to two groups receiving either coated or uncoated introducer sheaths. Radifocus Introducer II (Terumo) 25 cm, 6 Fr radial sheaths and sheaths that were identical apart from the presence of the coating were used in all patients. The APD was used for sheath removal at the end of the procedure. Three patients (7%) in the coated group experienced discomfort during automatic sheath removal, compared to 12 patients (27%) in the uncoated group (P = 0.02). The maximum pullback force (MPF) was significantly lower in the coated compared to the uncoated group (0.24 ± 0.31 vs. 0.44 ± 0.33 kg; P = 0.003). Similarly, the mean pullback force was significantly lower in the coated group (0.14 ± 0.23 vs. 0.32 ± 0.24 kg; P < 0.001). Only one patient (2%) in each group had an MPF greater than 1.0 kg together with clinical evidence of radial artery spasm. Removal of the coated Terumo Radifocus sheath requires less force than an identical uncoated sheath. The coated sheath was also associated with less discomfort for the patient. Cathet Cardiovasc Intervent 2003;59:161–164.Keywords
This publication has 11 references indexed in Scilit:
- Evaluation of a spasmolytic cocktail to prevent radial artery spasm during coronary proceduresCatheterization and Cardiovascular Interventions, 2003
- Measurement of radial artery spasm using an automatic pullback device.Catheterization and Cardiovascular Interventions, 2001
- Reduction of discomfort at sheath removal during transradial coronary procedures with the use of a hydrophilic-coated sheathCatheterization and Cardiovascular Interventions, 2001
- Stenting in acute coronary syndromes: a comparison of radial versus femoral access sitesJournal of the American College of Cardiology, 1998
- Coronary angiography from the radial artery – experience, complications and limitationsInternational Journal of Cardiology, 1998
- Characteristics of adrenoceptors in the human radial artery: Clinical implicationsThe Journal of Thoracic and Cardiovascular Surgery, 1998
- A Randomized Comparison of Percutaneous Transluminal Coronary Angioplasty by the Radial, Brachial and Femoral Approaches: The Access StudyJournal of the American College of Cardiology, 1997
- Outpatient Coronary Stent ImplantationJournal of the American College of Cardiology, 1997
- Comparison among arterial grafts and coronary artery: An attempt at functional classificationThe Journal of Thoracic and Cardiovascular Surgery, 1995
- Transradial artery coronary angioplastyAmerican Heart Journal, 1995