“Motion-Resistant” Pulse Oximetry: A Comparison of New and Old Models
- 1 October 2002
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 95 (4) , 967-972
- https://doi.org/10.1097/00000539-200210000-00033
Abstract
Several pulse oximeter manufacturers have recently developed instruments that are claimed to be resistant to the effects of patient motion. We performed a laboratory volunteer experiment to compare the performances of several of these instruments, as well as some older models, during combinations of motion and hypoxemia. Twenty oximeters were studied. A motorized table produced different hand motions, and each motion was studied during both room air breathing and hypoxemia. Pulse oximeters on the nonmoving hand were used to provide control measurements for comparison. The Masimo SET((R)) pulse oximeter exhibited the best overall performance, with a performance index (percentage of time in which the SpO(2) reading is within 7% of control value) of 94%. The Agilent Viridia 24C was next, with an 84% index, followed by the Agilent CMS (80%), the Datex-Ohmeda 3740 (80%), and the Nellcor N-395 (69%). For comparison with older oximeter technology, the Criticare 5040 had an index of 28%. Recent technology changes have significantly improved pulse oximeter performance during motion artifact, with the Masimo oximeter leading the way. IMPLICATIONS. New improvements in pulse oximeter technology have resulted in significantly better accuracy and reliability during patient motion. The Masimo pulse oximeter demonstrated the best performance of the 20 instruments tested.Keywords
This publication has 13 references indexed in Scilit:
- Room 301, 10/16/2000 9: 00 AM - 10: 30 AM (PD) Sensitivity and Specificity Performance during Motion Artifact in Three Pulse Oximeters Designed for Use in MotionAnesthesiology, 2000
- Comparison of the Nellcor N‐200 and N‐3000 pulse oximeters during simulated postoperative activitiesAnaesthesia, 1997
- The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)Anesthesiology, 1997
- Identification of movement artefact by the Nellcor N-200 and N-3000 pulse oximetersJournal of Clinical Monitoring and Computing, 1997
- Evaluation of the influence of movement on saturation readings from pulse oximetersAnaesthesia, 1995
- Crying wolfCritical Care Medicine, 1994
- Randomized Evaluation of Pulse Oximetry in 20,802 Patients; IAnesthesiology, 1993
- A prospective study of intraoperative pulse oximetry failureJournal of Clinical Monitoring and Computing, 1991
- EFFECT OF MOTION ARTEFACT ON PULSE OXIMETERS: EVALUATION OF FOUR INSTRUMENTS AND FINGER PROBESBritish Journal of Anaesthesia, 1990
- Measurement in Medicine: The Analysis of Method Comparison StudiesJournal of the Royal Statistical Society: Series D (The Statistician), 1983