Comparison of nine commercial immunoassays for the detection of rotavirus in fecal specimens
- 1 September 1988
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 26 (9) , 1630-1634
- https://doi.org/10.1128/jcm.26.9.1630-1634.1988
Abstract
One hundred fecal specimens obtained from patients with acute gastroenteritis were tested for rotavirus with nine commercial immunoassays to evaluate the sensitivity, specificity, predictive value, and diagnostic accuracy of these assays. Kits evaluated included two monoclonal antibody-based enzyme immunoassays (EIAs) (Rotaclone and Pathfinder Rotavirus), three polyclonal antibody-based EIAs (Rotavirus Immunoassay, Rotazyme II, and Wellcozyme Rotavirus), and four latex agglutinatin assays (Rotastat, Virogen Rotatest, Meritec-Rotavirus, and The Wellcome Latex Test). Thirty-eight of the 100 specimens were found to contain rotavirus by a reference microplate EIA. The accuracy of the reference assay was determined by RNA electrophoresis and a blocking assay on discordant specimens. The two monoclonal antibody EIAs had superior sensitivities (100%) and identified two positive specimens which were negative by the reference method but positive by the blocking assay. Among the polyclonal EIAs, all had sensitivities of greater than 90%, but specificities were variable; Rotazyme II, with a specificity of 50%, showed considerable discrepancy from other polyclonal EIAs. The latex tests had sensitivities ranging from 70 to 90% and specificities of 80 to 100%. Latex agglutination tests were more rapid than EIAs and did not require expensive equipment. The final choice of assay system will depend on the cost, speed, and accuracy requirements of the clinical laboratory.This publication has 40 references indexed in Scilit:
- Comparison of seven kits for detection of rotavirus in fecal specimens with a sensitive, specific enzyme immunoassayDiagnostic Microbiology and Infectious Disease, 1987
- Comparison of six methods for detecting human rotavirus in stoolsEuropean Journal of Clinical Microbiology & Infectious Diseases, 1984
- A ROTAVIRUS STAPHYLOCOCCAL CO‐AGGLUTINATION TESTActa Pathologica Microbiologica Scandinavica Series B: Microbiology, 1983
- Unreliability of Rotazyme ELISA test in neonatesThe Journal of Pediatrics, 1983
- Prevalence of rotavirus in children in day care centersThe Journal of Pediatrics, 1983
- Rotavirus: A cause of nosocomial infection in the nurseryThe Journal of Pediatrics, 1982
- Isolation of human rotavirus in cell culturesArchiv für die gesamte Virusforschung, 1981
- Reovirus-like agent as a cause of nosocomial diarrhea in infantsThe Journal of Pediatrics, 1977
- Diagnostic electron microscopy of faecesJournal of Clinical Pathology, 1974
- Cleavage of Structural Proteins during the Assembly of the Head of Bacteriophage T4Nature, 1970