Stability of Pacing Threshold, Impedance, and R Wave Amplitude at Rest and During Exercise
- 1 December 1990
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 13 (12) , 1602-1608
- https://doi.org/10.1111/j.1540-8159.1990.tb06861.x
Abstract
The pacing threshold of the human heart may be altered by physiological factors such as physical exercise. These changes may influence the individual programming of a pacemaker, since pacemakers can be programmed at pulse amplitudes of 2.5 volts and less. We investigated 22 patients with a multiprogrammable ventricular demand pacemaker 3 months after implantation; 16 patients had received a steroid-eluting lead and six patients had an Elgiloy lead. Parameters measured at rest and immediately after exercise were: Voltage threshold at pulse durations between 0.05 and 0.6 ms, impedance, R wave amplitude and energy consumption for the pacing threshold at 0.5 ms pulse duration. All patients performed a symptom limited supine bicycle exercise test. None of the investigated parameters showed a significant difference between rest and exercise, neither for the steroid eluting lead nor for the Elgiloy lead. The data suggest that the individual programming of a pacemaker adapted to the measurements at rest is also reliable and safe during exercise.Keywords
This publication has 12 references indexed in Scilit:
- A Stagewise Rejective Multiple Test Procedure Based on a Modified Bonferroni TestBiometrika, 1988
- Decrease in Canine Endocardial and Epicardial Electrogram Voltages with Exercise: Implications for Pacemaker SensingPacing and Clinical Electrophysiology, 1988
- A stagewise rejective multiple test procedure based on a modified Bonferroni testBiometrika, 1988
- Clinical Results of Low Energy Unipolar or Bipolar Activated Carbon Tip LeadsPacing and Clinical Electrophysiology, 1986
- Clinical Experience with 910 Carbon Tip Leads: Comparison with Polished Platinum Leads.Pacing and Clinical Electrophysiology, 1986
- Long‐Term Performance of Endocardial Leads with Steroid‐Eluting ElectrodesPacing and Clinical Electrophysiology, 1986
- Acute and Long‐Term Atrial and Ventricular Stimulation Thresholds with a Steroid‐Eluting ElectrodePacing and Clinical Electrophysiology, 1985
- Stimulation and sensing thresholds for cardiac pacing: Electrophysiologic and technical aspectsProgress in Cardiovascular Diseases, 1981
- PHYSIOLOGICAL CHANGES IN THRESHOLDAnnals of the New York Academy of Sciences, 1969
- Changes in myocardial threshold. Physiologic and pharmacologic factors in patients with implanted pacemakersAmerican Heart Journal, 1967