Transthoracic Cardioversion of Atrial Fibrillation
- 21 March 2000
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Circulation
- Vol. 101 (11) , 1282-1287
- https://doi.org/10.1161/01.cir.101.11.1282
Abstract
Clinical studies have shown that biphasic shocks are more effective than monophasic shocks for ventricular defibrillation. The purpose of this study was to compare the efficacy of a rectilinear biphasic waveform with a standard damped sine wave monophasic waveform for the transthoracic cardioversion of atrial fibrillation. In this prospective, randomized, multicenter trial, patients undergoing transthoracic cardioversion of atrial fibrillation were randomized to receive either damped sine wave monophasic or rectilinear biphasic shocks. Patients randomized to the monophasic protocol (n=77) received sequential shocks of 100, 200, 300, and 360 J. Patients randomized to the biphasic protocol (n=88) received sequential shocks of 70, 120, 150, and 170 J. First-shock efficacy with the 70-J biphasic waveform (60 of 88 patients, 68%) was significantly greater than that with the 100-J monophasic waveform (16 of 77 patients, 21%, P<0.0001), and it was achieved with 50% less delivered current (11+/-1 versus 22+/-4 A, P<0.0001). Similarly, the cumulative efficacy with the biphasic waveform (83 of 88 patients, 94%) was significantly greater than that with the monophasic waveform (61 of 77 patients, 79%; P=0.005). The following 3 variables were independently associated with successful cardioversion: use of a biphasic waveform (relative risk, 4.2; 95% confidence intervals, 1.3 to 13.9; P=0.02), transthoracic impedance (relative risk, 0.64 per 10-Omega increase in impedance; 95% confidence intervals, 0.46 to 0.90; P=0.005), and duration of atrial fibrillation (relative risk, 0.97 per 30 days of atrial fibrillation; 95% confidence intervals, 0.96 to 0.99; P=0.02). For transthoracic cardioversion of atrial fibrillation, rectilinear biphasic shocks have greater efficacy (and require less energy) than damped sine wave monophasic shocks.Keywords
This publication has 19 references indexed in Scilit:
- Comparison of a novel rectilinear biphasic waveform with a damped sine wave monophasic waveform for transthoracic ventricular defibrillationJournal of the American College of Cardiology, 1999
- Comparison of biphasic and monophasic shocks for defibrillation using a nonthoracotomy systemThe American Journal of Cardiology, 1993
- Comparison of the internal defibrillation thresholds for monophasic and double and single capacitor biphasic waveformsJournal of the American College of Cardiology, 1989
- A prospective randomized evaluation of biphasic versus monophasic waveform pulses on defibrillation efficacy in humansJournal of the American College of Cardiology, 1989
- Improved internal defibrillation efficacy with a biphasic waveformAmerican Heart Journal, 1989
- Improved low energy defibrillation efficacy in man with the use of a biphasic truncated exponential waveformAmerican Heart Journal, 1989
- Current-based versus energy-based ventricular defibrillation: A prospective studyJournal of the American College of Cardiology, 1988
- Relationship between canine transthoracic impedance and defibrillation threshold. Evidence for current-based defibrillation.Journal of Clinical Investigation, 1987
- Self-adhesive preapplied electrode pads for defibrillation and cardioversionJournal of the American College of Cardiology, 1984
- Elective Cardioversion: Influence of Paddle-Electrode Location and Size on Success Rates and Energy RequirementsNew England Journal of Medicine, 1981