Characteristics of the m 2000 Automated Sample Preparation and Multiplex Real-Time PCR System for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae
- 1 March 2007
- journal article
- research article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 45 (3) , 747-751
- https://doi.org/10.1128/jcm.01956-06
Abstract
We evaluated a new real-time PCR-based prototype assay for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae developed by Abbott Molecular Inc. This assay is designed to be performed on an Abbott m2000 real-time instrument system, which consists of an m2000sp instrument for sample preparation and an m2000rt instrument for real-time PCR amplification and detection. The limit of detection of this prototype assay was determined to be 20 copies of target DNA for both C. trachomatis and N. gonorrhoeae, using serially diluted linearized plasmids. No cross-reactivity could be detected when 55 nongonococcal Neisseria isolates and 3 non-C. trachomatis Chlamydia isolates were tested at 1 million genome equivalents per reaction. Concordance with the Roche Amplicor, BDProbeTec ET, and Gen-Probe APTIMA Combo 2 tests was assessed using unlinked/deidentified surplus clinical specimens previously analyzed with these tests. For C. trachomatis, concordance for positive results ranged from 93.7% to 100%, while concordance for negative results ranged from 98.2% to 100%. For N. gonorrhoeae, concordance for positive and negative results ranged from 91.4% to 100% and 99.3% to 100%, respectively. A workflow analysis of the prototype assay was conducted to obtain information on throughput under laboratory conditions. At 48 samples/run, the time to first result for both C. trachomatis and N. gonorrhoeae was 4.5 h. A total of 135 patient specimens could be analyzed in 8.9 h, with 75 min of hands-on time. This study demonstrated the technical and clinical feasibility of the new Abbott real-time PCR C. trachomatis/N. gonorrhoeae assay.Keywords
This publication has 26 references indexed in Scilit:
- High Analytical Sensitivity and Low Rates of Inhibition May Contribute to Detection of Chlamydia trachomatis in Significantly More Women by the APTIMA Combo 2 AssayJournal of Clinical Microbiology, 2006
- Detection of Chlamydia trachomatis and Neisseria gonorrhoeae by Strand Displacement Amplification and Relevance of the Amplification Control for Use with Vaginal Swab SpecimensJournal of Clinical Microbiology, 2003
- Vaginal Swabs Are Appropriate Specimens for Diagnosis of Genital Tract Infection with Chlamydia trachomatisJournal of Clinical Microbiology, 2003
- Reproducibility of Positive Test Results in the BDProbeTec ET System for Detection of Chlamydia trachomatis and Neisseria gonorrhoeaeJournal of Clinical Microbiology, 2003
- Specimen Processing and Concentration of Chlamydia trachomatis Added Can Influence False-Negative Rates in the LCx Assay but Not in the APTIMA Combo 2 Assay When Testing for InhibitorsJournal of Clinical Microbiology, 2003
- Evaluation of an Automated Liquid-Handling System (Tecan Genesis RSP 100) in the Abbott LCx Assay for Chlamydia trachomatisJournal of Clinical Microbiology, 2001
- Detection of Chlamydia trachomatis and Neisseria gonorrhoeae by Enzyme Immunoassay, Culture, and Three Nucleic Acid Amplification TestsJournal of Clinical Microbiology, 2001
- Multicenter Evaluation of the BDProbeTec ET System for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae in Urine Specimens, Female Endocervical Swabs, and Male Urethral SwabsJournal of Clinical Microbiology, 2001
- VARIATION AND CONTROL OF PROTEIN EXPRESSION IN NEISSERIAAnnual Review of Microbiology, 1990
- A common plasmid of Chlamydia trachomatisPlasmid, 1986