Reproducibility of Positive Test Results in the BDProbeTec ET System for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae
Open Access
- 1 August 2003
- journal article
- Published by American Society for Microbiology in Journal of Clinical Microbiology
- Vol. 41 (8) , 3911-3914
- https://doi.org/10.1128/jcm.41.8.3911-3914.2003
Abstract
Nucleic acid amplification tests such as the BDProbeTec ET (BDPT) system are more prone to reproducibility problems than are antigen detection tests and culture. A repeat testing algorithm for all samples with method other than acceleration (MOTA) scores greater than or equal to the cutoff value (2,000) was developed for the BDPT system and applied in a clinical laboratory setting. All positive samples were retested, and if the result of the second test was below the cutoff value, a third test was performed to resolve the discrepancy. Overall, 11 (5.3%) of 207 samples initially positive for Chlamydia trachomatis and 11 (10.7%) of 103 samples initially positive for Neisseria gonorrhoeae were not confirmed by repeat testing of the original sample. Poor reproducibility was associated with low-positive MOTA scores (2,000 to 9,999) for both analytes. Only 21 (80.8%) of 26 low-positive samples in the C. trachomatis test and 4 (33.3%) of 12 low-positive samples in the N. gonorrhoeae test retested as positive. The reproducibility of both tests with samples with initial MOTA scores of ≥10,000 increased to 96.7%. The data suggest that retesting of low-positive samples is warranted and could reduce the number of potentially false-positive test results.Keywords
This publication has 9 references indexed in Scilit:
- Screening tests to detect Chlamydia trachomatis and Neisseria gonorrhoeae infections--2002.2002
- Accuracy of Results Obtained by Performing a Second Ligase Chain Reaction Assay and PCR Analysis on Urine Samples with Positive or Near-Cutoff Results in the LCx Test for Chlamydia trachomatisJournal of Clinical Microbiology, 2002
- Multicenter Evaluation of the BDProbeTec ET System for Detection of Chlamydia trachomatis and Neisseria gonorrhoeae in Urine Specimens, Female Endocervical Swabs, and Male Urethral SwabsJournal of Clinical Microbiology, 2001
- Reproducibility problems with the Abbott laboratories LCx assay for Chlamydia trachomatis and Neisseria gonorrhoeae.2000
- Reproducibility Problems with the Abbott Laboratories LCx Assay for Chlamydia trachomatis and Neisseria gonorrhoeaeJournal of Clinical Microbiology, 2000
- Strand Displacement Amplification and Homogeneous Real-Time Detection Incorporated in a Second-Generation DNA Probe System, BDProbeTecETClinical Chemistry, 1999
- Reproducibility of the Roche Amplicor Polymerase Chain Reaction Assay for Detection of Infection by Chlamydia trachomatis in Endocervical SpecimensClinical Chemistry, 1998
- Reproducibility problems with the AMPLICOR PCR Chlamydia trachomatis testJournal of Clinical Microbiology, 1997