Penetration Effects in Wide Aperture Collimators
- 1 July 1968
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 13 (3) , 401-412
- https://doi.org/10.1088/0031-9155/13/3/307
Abstract
The problem of penetration of y-rays through wide aperture cylindrical collimators, placed close to a body organ for radioisotope uptake studies, is examined. Values for the penetration factors are numerically computed for axial point sources and compared with those given by the radius-independent formulae, normally used, to establish criteria for the use of these formulae. Penetration factors are also measured experimentally, and within the limitations of the experimental method there is good agreement with the computed values, even for sources close to wide aperture collimators.Keywords
This publication has 3 references indexed in Scilit:
- The Calculation of Gamma Ray Penetration of the Walls of Cylindrical and Conical Collimating HolesPhysics in Medicine & Biology, 1962
- A Study of the External Gamma-ray Field Arising from a Radioactive Source Situated in Air and in a Scattering MediumPhysics in Medicine & Biology, 1959
- Gamma-Ray Collimator Penetration and Scattering EffectsJournal of Applied Physics, 1957