Quality Specifications for Glucose Meters: Assessment by Simulation Modeling of Errors in Insulin Dose
- 1 February 2001
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 47 (2) , 209-214
- https://doi.org/10.1093/clinchem/47.2.209
Abstract
Background: Proposed quality specifications for glucose meters allow results to be in error by 5–10% or more of the “true” concentration. Because meters are used as aids in the adjustment of insulin doses, we aimed to characterize the quantitative effect of meter error on the ability to identify the insulin dose appropriate for the true glucose concentration. Methods: Using Monte Carlo simulation, we generated random “true” glucose values within defined intervals. These values were converted to “measured” glucose values using mathematical models of glucose meters having defined imprecision (CV) and bias. For each combination of bias and imprecision, 10 000–20 000 true and measured glucose concentrations were matched with the corresponding insulin doses specified by selected insulin-dosing regimens. Discrepancies in prescribed doses were counted and their frequencies plotted in relation to bias and imprecision. Results: For meters with a total analytical error of 5%, dosage errors occurred in ∼8–23% of insulin doses. At 10% total error, 16–45% of doses were in error. Large errors of insulin dose (two-step or greater) occurred >5% of the time when the CV and/or bias exceeded 10–15%. Total dosage error rates were affected only slightly by choices of sliding scale among insulin dosage rules or by the range of blood glucose. To provide the intended insulin dosage 95% of the time required that both the bias and the CV of the glucose meter be Conclusions: Glucose meters that meet current quality specifications allow a large fraction of administered insulin doses to differ from the intended doses. The effects of such dosage errors on blood glucose and on patient outcomes require study.Keywords
This publication has 10 references indexed in Scilit:
- Newer Portable Glucose Meters—Analytical Improvement Compared with Previous Generation Devices?Clinical Chemistry, 1999
- The effects of exercise during hemodialysis on physical performance and nutrition assessmentJournal of Renal Nutrition, 1999
- Analytical Performance Characteristics Should Be Judged against Objective Quality SpecificationsClinical Chemistry, 1999
- Analytical Error of Home Glucose Monitors: A Comparison of 18 SystemsAnnals of Clinical Biochemistry: International Journal of Laboratory Medicine, 1999
- Sell-Monitoring of Blood GlucoseDiabetes Care, 1996
- Charts of operational process specifications ("OPSpecs charts") for assessing the precision, accuracy, and quality control needed to satisfy proficiency testing performance criteriaClinical Chemistry, 1992
- The Necessity of Achieving Good Laboratory PerformanceDiabetic Medicine, 1990
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- Comparison Between Continuous Subcutaneous Insulin Infusion and Multiple Injections of Insulin: A One-Year Prospective StudyDiabetes, 1982
- Biological and Analytic Components of Variation in Long-Term Studies of Serum Constituents in Normal SubjectsClinical Chemistry, 1970