A High-Resolution, Small Diameter Image Intensifier Gamma Camera System
- 1 April 1967
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 12 (2) , 251-255
- https://doi.org/10.1088/0031-9155/12/2/310
Abstract
A stationary gamma ray detector incorporating a light image intensifier is described. The light emitted from a 2 in. diameter, 3 mm thick, flat crystal is focused, by means of a special Perspex lens system, onto the photocathode of the image intensifier (also 2 in. diameter) with a 5% efficiency. Using a parallel-hole lead collimator 2 in. long and having 2 mm square holes and 0.007 in thick septa, it was found that, when using a 99mTc source (140 kev), a resolution of 1 mm was obtainable, although contrast was poorer at this value, due in part to the acceptance of unwanted Compton events within the subject, collimator and crystal. Contrast at separations of the order of 2 mm and above was quite good, however. A comparison with other image amplifier systems and with the Anger camera is made, and future plans are discussed.Keywords
This publication has 4 references indexed in Scilit:
- A New Method of Assessing the Performance of Scintillation Cameras and ScannersPhysics in Medicine & Biology, 1966
- The Resolution of a Spark Imaging Camera (Spintharicon) with Parallel Channel CollimatorsPhysics in Medicine & Biology, 1966
- An image intensifier-scintillator device for determination of profiles and images of weak beams of ionizing particles or X-raysNuclear Instruments and Methods, 1961