Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium
Top Cited Papers
- 1 October 2004
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 31 (10) , 2899-2911
- https://doi.org/10.1118/1.1788932
Abstract
An extensive set of benchmark measurement of PDDs and beam profiles was performed in a heterogeneous layer phantom, including a lung equivalent heterogeneity, by means of several detectors and compared against the predicted dose values by different calculation algorithms in two treatment planning systems. PDDs were measured with TLDs, plane parallel and cylindrical ionization chambers and beam profiles with films. Additionally, Monte Carlo simulations by meansof the PENELOPE code were performed. Four different field sizes (, , , and) and two lung equivalent materials (CIRS, and St. Bartholomew Hospital, London, ) were studied. The performance of four correction‐based algorithms and one based on convolution‐superposition was analyzed. The correction‐based algorithms were the Batho, the Modified Batho, and the Equivalent TAR implemented in the Cadplan (Varian) treatment planning system and the TMS Pencil Beam from the Helax‐TMS (Nucletron) treatment planning system. The convolution‐superposition algorithm was the Collapsed Cone implemented in the Helax‐TMS. The only studied calculation methods that correlated successfully with the measured values with a 2% average inside all media were the Collapsed Cone and the Monte Carlo simulation. The biggest difference between the predicted and the delivered dose in the beam axis was found for the EqTAR algorithm inside the CIRS lung equivalent material in a x‐ray beam. In these conditions, average and maximum difference against the TLD measurements were 32% and 39%, respectively. In the water equivalent part of the phantom every algorithm correctly predicted the dose (within 2%) everywhere except very close to the interfaces where differences up to 24% were found for photon beams. Consistent values were found between the reference detector (ionization chamber in water and TLD in lung) and Monte Carlo simulations, yielding minimal differences . The penumbra broadening effect in low density media was not predicted by any of the correction‐based algorithms, and the only one that matched the experimental values and the Monte Carlo simulations within the estimated uncertainties was the Collapsed Cone Algorithm.Keywords
This publication has 36 references indexed in Scilit:
- Experimental validation tests of fast Fourier transform convolution and multigrid superposition algorithms for dose calculation in low-density mediaRadiotherapy and Oncology, 2003
- Effect of low-density lateral interfaces on soft-tissue dosesInternational Journal of Radiation Oncology*Biology*Physics, 1997
- Comparison of dose homogeneity effects due to electron equilibrium loss in lung for 6 MV and 18 MV photonsInternational Journal of Radiation Oncology*Biology*Physics, 1996
- The effect of secondary electron spread on the penumbra in high energy photon beam therapyThe British Journal of Radiology, 1990
- Evaluation of lung dose correction methods for photon irradiations of thorax phantomsInternational Journal of Radiation Oncology*Biology*Physics, 1989
- Benchmark measurements for lung dose corrections for x-ray beamsInternational Journal of Radiation Oncology*Biology*Physics, 1988
- A method for conversion of hounsfield number to electron density and prediction of macroscopic pair production cross-sectionsRadiotherapy and Oncology, 1986
- Accuracy of lung dose calculations for large‐field irradiation with 6‐MV x raysMedical Physics, 1986
- Underdosing of lesions resulting from lack of electronic equilibrium in upper respiratory air cavities irradiated by 10 mv X-ray beamsInternational Journal of Radiation Oncology*Biology*Physics, 1977