Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube
Open Access
- 1 October 2003
- journal article
- research article
- Published by Springer Nature in Intensive Care Medicine
- Vol. 29 (10) , 1849-1853
- https://doi.org/10.1007/s00134-003-1933-6
Abstract
This study compared the recently introduced Microcuff endotracheal tube HVLP ICU featuring an ultrathin (7-µm) polyurethane cuff membrane with endotracheal tubes from different manufacturers regarding fluid leakage past the tube cuff. In vitro setup. The following endotracheal tubes (ID 7.5 mm) were compared: Mallinckrodt HiLo, Microcuff HVLP ICU, Portex Profile Soft Seal, Rüsch Super Safety Clear, and Sheridan CF. A vertical PVC trachea model (ID 20 mm) was intubated, and cuffs were inflated to 10, 15, 20, 25, 30, and 60 cmH2O. Colored water (5 ml) was added to the top of the cuff. The amount of leaked fluid past the tube cuff within 5, 10, and 60 min was recorded. Experiments were performed four times using two examples of each tube brand. Fluid leakage past tube cuffs occurred in all conventional endotracheal tubes at cuff pressures from 10 to 60 cmH2O. In the Microcuff tube cuff pressure fluid leakage was observed within 10 min only at 10 cmH2O. Results with the Microcuff tube were significantly better than all other tube brands at cuff pressures of 10–30 cmH2O. Within the acceptable upper limit for tracheal cuff pressure (25–30 cmH2O) the Microcuff endotracheal tube was the only one of the tested tubes to prevent fluid leakage in our in vitro setup. In vivo studies are required to confirm these findings.Keywords
This publication has 14 references indexed in Scilit:
- Tracheal cuff pressure--a survey of clinical practice.2002
- The margin of safety associated with the use of cuffed paediatric tracheal tubesAnaesthesia, 2002
- Gel Lubrication of the Tracheal Tube Cuff Reduces Pulmonary AspirationAnesthesiology, 2001
- Leakage of fluid around high‐volume, low‐pressure cuffsAnaesthesia, 2001
- Prevention of tracheal aspiration using the pressure‐limited tracheal tube cuffAnaesthesia, 1999
- Leakage of fluid around low‐pressure tracheal tube cuffsAnaesthesia, 1997
- Leakage of fluid past the tracheal tube cuff in a benchtop modelBritish Journal of Anaesthesia, 1997
- GASTRODUODENAL DYSFUNCTION AS A CAUSE OF GASTRIC BACTERIAL OVERGROWTH IN PATIENTS UNDERGOING MECHANICAL VENTILATION OF THE LUNGSBritish Journal of Anaesthesia, 1992
- Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs.BMJ, 1984
- THE CROSS-SECTIONAL SHAPE AND CIRCUMFERENCE OF THE HUMAN TRACHEA1984