Boneloc® cemented total hip prostheses: Loosening in 28/43 cases after 3-38 months
- 1 January 1995
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 66 (3) , 215-217
- https://doi.org/10.3109/17453679508995526
Abstract
We report our early results with 43 total hip prostheses (25 Charnley and 18 LMT) inserted during 1991 with Boneloc® cement. The indication was primary arthrosis (38) and rheumatoid arthritis (5). After 18 (3-38) months, 28 stems were loose and 18 hips have been revised 1-4 years after primary surgery.Keywords
This publication has 10 references indexed in Scilit:
- Innovations in acrylic bone cement and application equipmentJournal of Applied Biomaterials, 1995
- Early failure with Boneloc® bone cement: 4/8 femoral stems loose within 3 yearsActa Orthopaedica, 1995
- Effects of polymerization heat and monomers from acrylic cement on canine boneActa Orthopaedica, 1994
- Improved cortical histology after cementation with a new MMA‐DMA‐IBMA bone cement: An animal studyJournal of Biomedical Materials Research, 1993
- The Use and Handling of Acrylic Bone Cement in Danish Orthopaedic DepartmentsBasic & Clinical Pharmacology & Toxicology, 1993
- Genotoxicity of Acrylic Bone CementsBasic & Clinical Pharmacology & Toxicology, 1991
- Post-traumatic cubitus varus. Long-term follow-up of corrective supracondylar humeral osteotomy in children.Journal of Bone and Joint Surgery, 1990
- Factors Associated with Early Loosening of Cemented Total Hip ProsthesesActa Orthopaedica, 1983