Assessment of Coronary Arterial Flow and Flow Reserve in Humans With Magnetic Resonance Imaging
- 15 April 1996
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 93 (8) , 1502-1508
- https://doi.org/10.1161/01.cir.93.8.1502
Abstract
Background The noninvasive measurement of absolute epicardial coronary arterial flow and flow reserve would be useful in the evaluation of patients with coronary circulatory disorders. Phase-contrast magnetic resonance imaging (PC-MRI) has been used to measure coronary arterial flow in animals, but its accuracy in humans is unknown. Methods and Results Twelve subjects (7 men, 5 women; age, 44 to 67 years) underwent PC-MRI measurements of flow in the left anterior descending coronary artery or one of its diagonal branches at rest and after administration of adenosine (140 μg·kg−1·min−1 IV). Immediately thereafter, intracoronary Doppler velocity (IDV) and flow measurements were made during cardiac catheterization at rest and after intravenous administration of adenosine. For the 12 patients, the correlation between MRI and invasive measurements of coronary arterial flow and coronary arterial flow reserve was excellent: coronary flowMRI (mL/min)= 0.85×coronary flowIDV (mL/min)+17 (mL/min), r=.89, and coronary flow reserveMRI=0.79×coronary velocity reserveIDV+0.34, r=.89. For the range of coronary arterial flows (18 to 161 mL/min) measured by MRI, the limit of agreement between MRI and catheterization measurements of flow was −13±30 mL/min; for the range of coronary reserves (0.7 to 3.7) measured by MRI, the limit of agreement between the two techniques was 0.1±0.4. Conclusions Cine velocity-encoded PC-MRI can noninvasively measure absolute coronary arterial flow in the left anterior descending artery in humans. PC-MRI can detect pharmacologically induced changes in coronary arterial flow and can reliably distinguish between those subjects with normal and abnormal coronary artery flow reserve.Keywords
This publication has 22 references indexed in Scilit:
- An improved quadrature or phased‐array coil for MR cardiac imagingMagnetic Resonance in Medicine, 1995
- In Vivo Validation of Magnetic Resonance Blood Volume Flow Measurements with Limited Spatial Resolution in Small VesselsMagnetic Resonance in Medicine, 1995
- Flow velocity quantification in human coronary arteries with fast, breath‐hold MR angiographyJournal of Magnetic Resonance Imaging, 1993
- Intracoronary pressure and flow velocity with sensor-tip guidewires: A new methodologic approach for assessment of coronary hemodynamics before and after coronary interventionsThe American Journal of Cardiology, 1993
- Assessment of angiographically intermediate coronary artery stenosis using the Doppler flowireThe American Journal of Cardiology, 1993
- Simultaneous measurement of coronary reserve flow by left anterior descending coronary artery doppler and great cardiac vein thermodilution methodsJournal of the American College of Cardiology, 1992
- Comparison of adenosine and exercise thallium-201 single-photon emission computed tomography (SPECT) myocardial perfusion imagingJournal of the American College of Cardiology, 1992
- In Vivo Validation of MR Velocity ImagingJournal of Computer Assisted Tomography, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Does Visual Interpretation of the Coronary Arteriogram Predict the Physiologic Importance of a Coronary Stenosis?New England Journal of Medicine, 1984