Comparison of the Evaporative Light Scattering Detector (ELSD) and Refractive Index Detector (RID) in Lipid Analysis
- 1 August 1993
- journal article
- research article
- Published by Taylor & Francis in Journal of Liquid Chromatography
- Vol. 16 (12) , 2469-2482
- https://doi.org/10.1080/10826079308019586
Abstract
The properties of an evaporative light scattering detector (ELSD) and a refractive index detector (RID) in isocratic HPLC analysis of lipids are compared. The effect of molecular structure, especially the unsaturation level, on the detector response is studied using eight different lipid standards. Linearity and sensitivity are compared. Quantitation of an unknown sample is also performed. The sensitivity of the ELSD was slightly better than that of the RID, the detector limits of triolein being 30 ng/injection and 50 ng/injection, respectively. The linearity of the ELSD was considered only approximate compared with that of the RID the response factors of the calibration curves being 0.9554 and 0.9999, respectively. The effect of the molecular structure on the detector response was higher with the RID than with the ELSD. The relative standard deviation (SD%) of responses was 23% for the RID and 12% for the ELSD. The unsaturation level had a greater effect on the detector response of the RID than on that of the ELSD. This study indicates that a standard deviation (SD) of at least 13% is to be expected with the ELSD and an even higher SD with the RI detector when quantitating heterogeneous mixtures or unknown compounds.Keywords
This publication has 19 references indexed in Scilit:
- Simulation of the mass response of the evaporative light scattering detectorAnalytical Chemistry, 1992
- Laser-light-scattering detection for high-speed countercurrent chromatographyJournal of Chromatography A, 1991
- Comparison of Detectors for Size Exclusion Chromatography of Heavy Oil Related SamplesJournal of Chromatographic Science, 1988
- Simple and rapid method for high-performance liquid chromatographic separation and quantification of soybean phospholipidsJournal of Chromatography A, 1988
- Analysis of Lipid Classes by HPLC with the Evaporative Light Scattering DetectorJournal of Liquid Chromatography, 1987
- Separation of lipid classes by high-performance liquid chromatography with the “mass detector”Journal of Chromatography A, 1986
- Study of the qualitative and quantitative properties of the light-scattering detectorJournal of Chromatography A, 1984
- Use of light scattering as a detector principle in liquid chromatographyJournal of Chromatography A, 1983
- The mass detector: A new detection system for carbohydrate and lipid analysesChromatographia, 1982
- Evaporative analyzer as a mass detector for liquid chromatographyAnalytical Chemistry, 1978