Correction factors for parallel-plate chambers used in plastic phantoms in electron dosimetry
- 1 November 1997
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 42 (11) , 2101-2118
- https://doi.org/10.1088/0031-9155/42/11/007
Abstract
In electron beam dosimetry using parallel-plate chambers solid phantoms are sometimes necessary. To obtain the dose to water from the ionization obtained in the solid phantom, fluence correction factors and perturbation factors have to be applied. In this study fluence factors in a perturbation free geometry have been determined experimentally for common phantom materials. Wall perturbation factors for simulated Attix, NACP, and Roos chambers have also been determined for the same materials. Comparative Monte Carlo calculations have been performed using the EGS4 Monte Carlo code. Comparison with data in newly published protocols such as IAEA and IPEMB shows an agreement with the results obtained in this paper to within 1%, demonstrating that the data published in these protocols may be used with reasonable accuracy if recommended phantoms are used. The results also show that if unsuitable phantom materials are used, the wall perturbation factors may differ for different chambers and for different phantom materials by more than 3% and perturbation factors have to be considered in order to obtain a high accuracy in the dose determination.Keywords
This publication has 16 references indexed in Scilit:
- Electron fluence correction factors for conversion of dose in plastic to dose in waterMedical Physics, 1997
- The IPEMB code of practice for electron dosimetry for radiotherapy beams of initial energy from 2 to 50 MeV based on an air kerma calibrationPhysics in Medicine & Biology, 1996
- Detour factors in water and plastic phantoms and their use for range and depth scaling in electron-beam dosimetryPhysics in Medicine & Biology, 1996
- Implication of electron backscattering for electron dosimetryPhysics in Medicine & Biology, 1991
- Electron backscatter corrections for parallel‐plate chambersMedical Physics, 1988
- Presta: The parameter reduced electron-step transport algorithm for electron monte carlo transportNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 1986
- Comparison of ionisation measurements in water and polystyrene for electron beam dosimetryPhysics in Medicine & Biology, 1985
- Dose errors due to charge storage in electron irradiated plastic phantomsMedical Physics, 1984
- Penetration of backscattered electrons in polystyrene for energies between 1 and 25 MeVPhysics in Medicine & Biology, 1982
- Einfluß der Vielfachstreuung von Elektronen auf die Ionisation in gasgefüllten HohlräumenRadiation and Environmental Biophysics, 1968