On the Use of Roentgen-Calibrated, Cavity-Ionization Chambers in Tissue-like Phantoms to determine Absorbed Dose
- 1 October 1962
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 7 (2) , 229-240
- https://doi.org/10.1088/0031-9155/7/2/309
Abstract
Roentgen-calibrated cavity-ionization chambers are used to determine point absorbed dose in a tissue-like phantom. A calibration factor for 2 MV x-rays includes corrections for wall attenuation of the chamber, for any lack of electronic equilibrium, and for the stopping power of the wall material relative to air. A further correction is required to determine the exposure dose that would have been obtained at the point in the absence of the chamber and any wall-thickening cap used in the calibration. The correction factor required to compensate for the attenuation of the phantom material removed to accommodate the chamber and its cap can be determined from experimental data already available. A rad/ roentgen conversion factor for 2 MV x-rays and a water phantom is calculated.Keywords
This publication has 9 references indexed in Scilit:
- Exposure Dose Measurements in Megavoltage TherapyPhysics in Medicine & Biology, 1962
- On the Use of Thimble Chambers in PhantomsRadiology, 1962
- The Stopping-power Correction for Graphite Cavity-chambers used with 2 mV X-raysPhysics in Medicine & Biology, 1959
- A Study of Cavity Ion Chambers for Use with 2 MV X-rays: Equilibrium Wall Thickness: Wall-Absorption CorrectionPhysics in Medicine & Biology, 1959
- The Present Status of Calibrations of Ionization Dosemeters with High-Voltage X-RaysPhysics in Medicine & Biology, 1956
- DEPTH DOSE DATA AND ISODOSE DISTRIBUTIONS FOR RADIATION FROM A 22 MEV BETATRON1949
- Radiation Dosimetry. Part IIThe British Journal of Radiology, 1937