Physical evaluation of computed radiography as a mammographic X-ray imaging system
- 1 October 1994
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 67 (802) , 988-996
- https://doi.org/10.1259/0007-1285-67-802-988
Abstract
The physical imaging properties of a computed radiography (CR) system operating under mammographic exposure conditions have been measured. These measurements include modulation transfer function (MTF), sensitometric response and noise power spectrum (NPS). These figures were used to derive signal-to-noise ratio (SNR) descriptors of the system performance. The same measures were derived for a mammographic film-screen system. The SNR properties of the CR system indicate superior low frequency performance over a wider dynamic range than the film-screen system. The spatial frequency dependent SNR properties of the CR system are, however, limited. The results of a psychophysical test support the results obtained from physical measurements. These results indicate that despite its limited sharpness, details of dimensions relevant to diagnosis in mammography (circular details 0.25 mm diameter to represent microcalcifications and 3 mm diameter to represent small masses) are reproduced using CR with SNRs which are at least comparable with those of a mammographic film-screen system.Keywords
This publication has 19 references indexed in Scilit:
- A clinical comparison between conventional and digital mammography utilizing computed radiographyThe British Journal of Radiology, 1994
- Signal, noise and SNR transfer properties of computed radiographyPhysics in Medicine & Biology, 1993
- Spatial Resolution in Digital MammographyInvestigative Radiology, 1993
- Compensation for field non-uniformity on a mammographic X-ray unitThe British Journal of Radiology, 1993
- A preliminary investigation of the imaging performance of photostimulable phosphor computed radiography using a new design of mammographic quality control test objectThe British Journal of Radiology, 1992
- A physical image quality evaluation of a digital spot fluorography systemPhysics in Medicine & Biology, 1992
- Comparison of Digital and Conventional Mammography: A ROC Study of 270 MammogramsMedical Informatics, 1992
- Digital Luminescence MammographyActa Radiologica, 1991
- Digitized Film RadiographyInvestigative Radiology, 1989