Subcutaneous glucose predicts plasma glucose independent of insulin: implications for continuous monitoring
- 1 September 1999
- journal article
- research article
- Published by American Physiological Society in American Journal of Physiology-Endocrinology and Metabolism
- Vol. 277 (3) , E561-E571
- https://doi.org/10.1152/ajpendo.1999.277.3.e561
Abstract
The present study investigated the relationship between blood and subcutaneous interstitial fluid (ISF) glucose by employing an amperometric glucose sensor specifically developed for 3-day continuous glucose monitoring. The apparent sensor sensitivity and ISF glucose equilibration delay were estimated on separate days during hyperglycemic clamps in four dogs in which insulin was either suppressed with somatostatin, allowed to change, or increased with an exogenous infusion. A 2-h sensor “settling-in” period was allowed before the clamps. During insulin deficiency, the sensor sensitivity and ISF glucose delay were 0.23 ± 0.03 nA per mg/dl and 4.4 ± 0.8 min. Sensitivity was not affected by increases in endogenous (0.30 ± 0.04 vs. 0.28 ± 0.04 nA per mg/dl) or exogenous insulin (0.18 ± 0.01 vs. 0.16 ± 0.01 nA per mg/dl) nor was the delay (3.3 ± 1.2 vs. 5.7 ± 1.1 and 9.2 ± 2.6 vs. 12.3 ± 1.7 min; P > 0.05 for all). Sensor glucose accurately predicted plasma glucose without correcting for delays 0.9 for all), whereas for longer delays a digital corrective filter was used (r = 0.91 with filter). We conclude that the relationship between blood and ISF glucose is not affected by insulin and that delays in ISF glucose equilibration can be corrected with digital filters.Keywords
This publication has 31 references indexed in Scilit:
- Transendothelial insulin transport is not saturable in vivo. No evidence for a receptor-mediated process.Journal of Clinical Investigation, 1996
- Subcutaneous Glucose Concentration in Humans: Real estimation and continuous monitoringDiabetes Care, 1995
- Continuous in vivo monitoring in diabetes: the subcutaneous glucose concentrationActa Anaesthesiologica Scandinavica, 1995
- Glucose Concentration in Subcutaneous Extracellular SpaceDiabetes Care, 1993
- On Line Continuous Monitoring of Subcutaneous Tissue Glucose is Feasible by Combining Portable Glucosensor with MicrodialysisHormone and Metabolic Research, 1993
- In vivo evaluation of an electroenzymatic glucose sensor implanted in subcutaneous tissueBiosensors and Bioelectronics, 1992
- Subcutaneous glucose monitoring by means of electrochemical sensors: fiction or reality?Journal of Biomedical Engineering, 1992
- An electroenzymatic glucose sensor fabricated on a flexible substrateSensors and Actuators B: Chemical, 1991
- Glucose Concentration at Possible Sensor Tissue Implant SitesDiabetes Care, 1982
- ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERYAnnals of the New York Academy of Sciences, 1962