Evaluation of the Sensititre system for quantitative antimicrobial drug susceptibility testing: a collaborative study
Open Access
- 1 March 1980
- journal article
- research article
- Published by American Society for Microbiology in Antimicrobial Agents and Chemotherapy
- Vol. 17 (3) , 464-469
- https://doi.org/10.1128/aac.17.3.464
Abstract
This three-center collaborative study was conducted to evaluate samples of Sensititre antimicrobial microdilution panels (GIBCO/INVENEX). Sensititre minimum inhibitory concentrations of 27 bacterial isolates were compared with those obtained by a reference microdilution method. The Sensititre and microdilution minimum inhibitory concentrations were equivalent within +/- 1 dilution in 87.6% of the comparable test results. Intralaboratory reproducibilities of the Sensititre and microdilution endpoints were equivalent with 80.4 and 82.4%, respetively, of on-scale endpoints in absolute agreement. Sensititre was more reproducible among laboratories, with nearly a 10% greater agreement of triplicate results. The Sensititre microdilution test as evaluated gave results which were essentially equivalent to those obtained with a standardized microdilution method.This publication has 4 references indexed in Scilit:
- Determination of antibiotic sensitivities by the Sensititre system.Journal of Clinical Pathology, 1978
- Evaluation of the Micro-Media System for Quantitative Antimicrobial Drug Susceptibility Testing: a Collaborative StudyAntimicrobial Agents and Chemotherapy, 1978
- Antibiotic sensitivity testing. Report of an international collaborative study.1971
- A Microdilution Method for Antibiotic Susceptibility Testing: An EvaluationAmerican Journal of Clinical Pathology, 1970