Long‐Term Reliability of Single Lead Atrial Synchronous Pacing Systems Using Closely Spaced Atrial Dipoles: Five‐Year Experience
- 1 November 1994
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 17 (11) , 1865-1869
- https://doi.org/10.1111/j.1540-8159.1994.tb03764.x
Abstract
To assess the long-term capability of single atrioven ticular (AV) lead VDD pacing systems using close atrial dipoles to assure reliable atrial guided pacing, the safety and efficacy of 86 VDD units implanted in 73 patients at a single center since November 1988 was reviewed. All patients suffered from advanced AV block with normal sinoatrial function. Sixty five patients received a LEM/CCS Twinal 30/30S system, four patients received a Vitatron-Saphir system, and four patients received a Medtronic Thera VDR 8348 system. All patients underwent provocative tests in search of myopotential interference, and Holter recordings; in a group of patients who underwent pacemaker replacement a comparison was made between implant and replacement measurements. The mean follow-up duration was 27.3 months. A high percentage of successfully VDD paced patients and a low incidence of pacemaker malfunction, regularly solved by pacemaker reprogramming, was reported. Atrial signal amplitudes comparable to those measured at implant were found at replacement in all patients. These data support the long-term reliability of single AV lead VDD pacing systems with closely spaced atrial dipoles, as well as stable atrial sensing by floating bipolar atrial electrodes and effective atrial synchronous ventricular pacing over time.Keywords
This publication has 6 references indexed in Scilit:
- Single Lead VDD pacing: an updatePublished by Springer Nature ,1994
- ABSTRACT SESSION 22: PACEMAKERS; FUNCTION AND DYSFUNCTIONPacing and Clinical Electrophysiology, 1993
- Italian Multicenter Study on a Single Lead VDD Pacing System Using a Narrow Atrial Dipole SpacingPacing and Clinical Electrophysiology, 1992
- European Multicenter Prospective Follow‐Up Study of 1,002 Implants of a Single Lead VDD Pacing SystemPacing and Clinical Electrophysiology, 1991
- Experience and Implantation Techniques with a New Single‐Pass Lead VDD Pacing SystemPacing and Clinical Electrophysiology, 1990
- Single‐Lead VDD Pacing SystemPacing and Clinical Electrophysiology, 1990