Reproducibility of total cerebral blood flow measurements using phase contrast magnetic resonance imaging
- 26 June 2002
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 16 (1) , 1-5
- https://doi.org/10.1002/jmri.10133
Abstract
To evaluate reproducibility of total cerebral blood flow (CBF) measurements with phase contrast magnetic resonance imaging (pcMRI). We repeated total CBF measurements in 15 healthy volunteers with and without cardiac triggering, and with and without repositioning. In eight volunteers measurements were performed at two different occasions. In addition, measurement of flow in a phantom was performed to validate MR measurements. A difference of 40.4 ml/minute was found between CBF measurements performed with and without triggering (P < 0.05). For repeated triggered measurements, the coefficient of variation (CV) was 7.1%, and for nontriggered measurements 10.3%. For repeated measurements with repositioning, the CV was 7.1% with and 11.2% without triggering. Repeated measurements at different occasions showed a CV of 8.8%. Comparing measured with real flow in the phantom, the triggered differed 4.9% and the nontriggered 8.3%. The findings of this study demonstrate that pcMRI is a reliable method to measure total CBF in terms of both accuracy and reproducibility. J. Magn. Reson. Imaging 2002;16:1–5.Keywords
This publication has 13 references indexed in Scilit:
- Effect of age on cerebral blood flow: measurement with ungated two-dimensional phase-contrast MR angiography in 250 adults.Radiology, 1998
- Measuring blood flow by nontriggered 2D phase-contrast MR angiographyMagnetic Resonance Imaging, 1996
- Accuracy and precision of time-averaged flow as measured by nontriggered 2D phase-contrast MR angiography, a phantom evaluationMagnetic Resonance Imaging, 1995
- Comparison of cerebral artery blood flow measurements with gated cine and ungated phase‐contrast techniquesJournal of Magnetic Resonance Imaging, 1993
- Nontriggered magnetic resonance velocity measurement of the time‐average of pulsatile velocityMagnetic Resonance in Medicine, 1993
- Computer-controlled flow simulator for MR flow studiesJournal of Magnetic Resonance Imaging, 1992
- Determination of cerebral blood flow with a phase-contrast cine MR imaging technique: evaluation of normal subjects and patients with arteriovenous malformations.Radiology, 1992
- Computer-controlled positive displacement pump for physiological flow simulationMedical & Biological Engineering & Computing, 1991
- The measurement of time-averaged flow by magnetic resonance imaging using continuous acquisition in the carotid arteries and its comparison with Doppler ultrasoundClinical Physics and Physiological Measurement, 1990
- Measurement of blood flow by ultrasound: Accuracy and sources of errorUltrasound in Medicine & Biology, 1985