Abstract
We have constructed a simple, all solid‐state, time‐correlated single photon counting device for collecting decay profiles of chromophores attached to DNA fragments moving through a capillary tube filled with a sieving gel under the influence of an applied electric field (capillary electrophoresis). The major components of the instrument consist of an actively pulsed GaAlAs diode laser (λexcitation=780 nm; τp6 cps). The instrument possessed a response function of approximately 275 ps (full width at half‐maximum), adequate for measuring fluorescence lifetimes in the subnanosecond regime. To demonstrate the utility and the sensitivity of the instrument, dynamic measurements of fluorescence lifetimes for near‐IR dye‐labeled DNA fragments were measured during capillary electrophoresis for the identification of the dye‐labeled nucleotide bases via temporal discrimination. The results indicated that in a two‐dye experiment, in which only two of the four constituent bases which comprise DNA were labeled with unique fluorescent probes, the characteristic lifetime of the probe could be used to readily identify the terminal nucleotide base. Decay profiles were constructed from roughly 17 500 photoelectrons accumulated over a 2 s counting interval at a loading level of approximately 6.2×10−21 mol (3900 molecules) of DNA per electrophoretic band. The lifetimes of the two labeling dyes were determined to be 669 ps (±42 ps) and 528 ps (±68 ps).Keywords
This publication has 14 references indexed in Scilit:
- Online Fluorescence Lifetime Determinations in Capillary ElectrophoresisAnalytical Chemistry, 1995
- Ultrasensitive Near-IR Fluorescence Detection for Capillary Gel Electrophoresis and DNA Sequencing ApplicationsAnalytical Chemistry, 1995
- Error Analysis of Simple Algorithms for Determining Fluorescence Lifetimes in Ultradilute Dye SolutionsApplied Spectroscopy, 1994
- Three DNA sequencing methods using capillary gel electrophoresis and laser-induced fluorescenceAnalytical Chemistry, 1991
- A System for Rapid DNA Sequencing with Fluorescent Chain-Terminating DideoxynucleotidesScience, 1987
- DNA sequence analysis with a modified bacteriophage T7 DNA polymerase.Proceedings of the National Academy of Sciences, 1987
- Fluorescence detection in automated DNA sequence analysisNature, 1986
- Pulsed semiconductor laser fluorometry for lifetime measurementsAnalytical Chemistry, 1985
- Better estimates of exponential decay parametersThe Journal of Physical Chemistry, 1981
- DNA sequencing with chain-terminating inhibitorsProceedings of the National Academy of Sciences, 1977