Peak Capacity, Peak-Capacity Production Rate, and Boiling Point Resolution for Temperature-Programmed GC with Very High Programming Rates
- 23 December 1999
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 72 (3) , 591-598
- https://doi.org/10.1021/ac9911802
Abstract
Recent advances in column heating technology have made possible very fast linear temperature programming for high-speed gas chromatography. A fused-silica capillary column is contained in a tubular metal jacket, which is resistively heated by a precision power supply. With very rapid column heating, the rate of peak-capacity production is significantly enhanced, but the total peak capacity and the boiling-point resolution (minimum boiling-point difference required for the separation of two nonpolar compounds on a nonpolar column) are reduced relative to more conventional heating rates used with convection-oven instruments. As temperature-programming rates increase, elution temperatures also increase with the result that retention may become insignificant prior to elution. This results in inefficient utilization of the downstream end of the column and causes a loss in the rate of peak-capacity production. The rate of peak-capacity production is increased by the use of shorter columns and higher carrier gas velocities. With high programming rates (100−600 °C/min), column lengths of 6−12 m and average linear carrier gas velocities in the 100−150 cm/s range are satisfactory. In this study, the rate of peak-capacity production, the total peak capacity, and the boiling point resolution are determined for C10−C28n-alkanes using 6−18 m long columns, 50−200 cm/s average carrier gas velocities, and 60−600 °C/min programming rates. It was found that with a 6-meter-long, 0.25-mm i.d. column programmed at a rate of 600 °C/min, a maximum peak-capacity production rate of 6.1 peaks/s was obtained. A total peak capacity of about 75 peaks was produced in a 37-s long separation spanning a boiling-point range from n-C10 (174 °C) to n-C28 (432 °C).Keywords
This publication has 15 references indexed in Scilit:
- Temperature Programming for High-Speed GCAnalytical Chemistry, 1999
- Application of Isotope Dilution to Ion Trap Gas Chromatography/Mass SpectrometryAnalytical Chemistry, 1998
- Column Selectivity Programming and Fast Temperature Programming for High-Speed GC Analysis of Purgeable Organic CompoundsAnalytical Chemistry, 1998
- Peer Reviewed: High-speed gas chromatographyAnalytical Chemistry, 1998
- Compact gas chromatograph probe for gas chromatography/mass spectrometry utilizing resistively heated aluminum-clad capillary columnsAnalytical Chemistry, 1989
- Optimization of separation number in gas chromatography with fused silica capillary columns under temperature programmed conditionsAnalytical Chemistry, 1983
- Evaluation of capillary columns by separation number or plate numberJournal of Chromatography A, 1981
- Comments on TZ: Its temperature dependence and relationship to other measures of column efficiencyJournal of High Resolution Chromatography, 1980
- Effect of test temperature on separation numberJournal of High Resolution Chromatography, 1980
- Separation values and their utilization in column characterizationChromatographia, 1975