Liquid−Liquid−Liquid Microextraction for Sample Preparation of Biological Fluids Prior to Capillary Electrophoresis
- 8 June 1999
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 71 (14) , 2650-2656
- https://doi.org/10.1021/ac990055n
Abstract
Methamphetamine as a model compound was extracted from 2.5-mL aqueous samples adjusted to pH 13 (donor solution) through a thin phase of 1-octanol inside the pores of a polypropylene hollow fiber and finally into a 25-μL acidic acceptor solution inside the hollow fiber. Following this liquid−liquid−liquid microextraction (LLLME), the acceptor solutions were analyzed by capillary zone electrophoresis (CE). Extractions were performed in simple disposable devices each consisting of a conventional 4-mL sample vial, two needles for introduction and collection of the acceptor solution, and a 8-cm piece of a porous polypropylene hollow fiber. From 5 to 20 different samples were extracted in parallel for 45 min, providing a high sample capacity. Methamphetamine was preconcentrated by a factor of 75 from aqueous standard solutions, human urine, and human plasma utilizing 10-1 M HCl as the acceptor phase and 10-1 M NaOH in the donor solution. In addition to preconcentration, LLLME also served as a technique for sample cleanup since large molecules, acidic compounds, and neutral components were not extracted into the acceptor phase. Utilizing diphenhydramine hydrochloride as internal standard, repetitive extractions varied less than 5.2% RSD (n = 6), while the calibration curve for methamphetamine was linear within the range 20 ng/μL to 10 μg/mL (r = 0.9983). The detection limit of methamphetamine utilizing LLLME/CE was 5 ng/mL (S/N = 3) in both human urine and plasma.Keywords
This publication has 15 references indexed in Scilit:
- Solvent Microextraction with Simultaneous Back-Extraction for Sample Cleanup and Preconcentration: Quantitative ExtractionAnalytical Chemistry, 1998
- Rapid loading of large sample volumes, analyte cleanup, and modified moving boundary transient isotachophoresis conditions for membrane preconcentration-capillary electrophoresis in small diameter capillariesElectrophoresis, 1996
- Enhanced performance membrane preconcentration–capillary electrophoresis–mass spectrometry (mPC‐CE‐MS) in conjunction with transient isotachophoresis for analysis of peptide mixturesJournal of High Resolution Chromatography, 1995
- Improved on-line membrane preconcentration-capillary electrophoresis (mPC-CE)Journal of High Resolution Chromatography, 1995
- Selective On-Line Preconcentration of Proteins by Tandem Metal Chelate Capillaries-Capillary Zone ElectrophoresisJournal of Liquid Chromatography, 1993
- Preparation and evaluation of an on‐line preconcentrator for capillary electrophoresisJournal of Microcolumn Separations, 1993
- Study of in Vivo pyrazoloacridine metabolism by capillary electrophoresis using isotachophoresis preconcentration in non‐aqueous separation bufferJournal of High Resolution Chromatography, 1993
- Sample stacking of an extremely large injection volume in high-performance capillary electrophoresisAnalytical Chemistry, 1992
- The Use of a Concentration Step to Collect Urinary Components Separated by Capillary Electrophoresis and Further Characterization of Collected Analytes by Mass SpectrometryJournal of Liquid Chromatography, 1991
- On-line coupling of capillary isotachophoresis with capillary zone electrophoresisJournal of Chromatography A, 1990