Design of field-cycled magnetic resonance systems for small animal imaging
- 17 May 2006
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 51 (11) , 2825-2841
- https://doi.org/10.1088/0031-9155/51/11/010
Abstract
This paper presents a design study for a field-cycled magnetic resonance imaging (MRI) system directed at small animal imaging applications. A field- cycled MRI system is different from a conventional MRI system in that it uses two separate and dynamically controllable magnetic fields. A strong magnetic field is used to polarize the object, and a relatively weak magnetic field is used during signal acquisition. The potential benefits of field-cycled MRI are described. The theoretical dependences of field-cycled MRI performance on system design are introduced and investigated. Electromagnetic, mechanical and thermal performances of the system were considered in this design study. A system design for imaging 10 cm diameter objects is presented as an example, capable of producing high-duty-cycle polarizing magnetic fields of 0.5 T and readout magnetic fields corresponding to a proton Larmor frequency of 5 MHz. The specifications of the final design are presented along with its expected electromagnetic and thermal performance.Keywords
This publication has 15 references indexed in Scilit:
- Rapid polarizing field cycling in magnetic resonance imagingIEEE Transactions on Medical Imaging, 2005
- Spatially resolved measurements of hyperpolarized gas properties in the lung in vivo. Part I: Diffusion coefficientMagnetic Resonance in Medicine, 1999
- A readout magnet for prepolarized MRIMagnetic Resonance in Medicine, 1996
- Novel approaches to low‐cost MRIMagnetic Resonance in Medicine, 1993
- MR relaxometry imaging. Work in progress.Radiology, 1992
- Environmental optimization and shielding for nmr experiments and imaging in the earth's magnetic fieldMagnetic Resonance in Medicine, 1990
- A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1–100 MHz: Dependence on tissue type, NMR frequency, temperature, species, excision, and ageMedical Physics, 1984
- Nuclear magnetism of liquid systems in the earth field rangePhysics Reports, 1980
- The sensitivity of the zeugmatographic experiment involving human samplesJournal of Magnetic Resonance (1969), 1979
- Transient Nuclear Induction Signals Associated with Pure Quadrupole InteractionsPhysical Review B, 1954