Noninvasive Blood Glucose Assay by Near-Infrared Diffuse Reflectance Spectroscopy of the Human Inner Lip
- 1 July 1993
- journal article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 47 (7) , 875-881
- https://doi.org/10.1366/0003702934415354
Abstract
Near-infrared (NIR) spectra of the human inner lip were obtained by using a special optimized accessory for diffuse reflectance measurements. The partial-least squares (PLS) multivariate calibration algorithm was applied for linear regression of the spectral data between 9000 and 5500 cm−1 (Λ = 1.1–1.8 μm) against blood glucose concentrations determined by a standard clinical enzymatic method. Calibration experiments with a single person were carried out under varying conditions, as well as with a population of 133 different patients, with capillary and venous blood glucose concentration values provided. A genuine correlation between the blood glucose concentrations and the NIR-spectra can be proven. A time lag of about 10 min for the glucose concentration in the spectroscopically probed tissue volume vs. the capillary concentration can be estimated. Mean-square prediction errors obtained by cross-validation were in the range of 45 to 55 mg/dL. An analysis of different variance factors showed that the major contribution to the average prediction uncertainty was due to the reduced measurement reproducibility, i.e., variations in lip position and contact pressure. The results demonstrate the feasibility of using diffuse reflectance NIR-spectroscopy for the noninvasive measurement of blood glucose.Keywords
This publication has 10 references indexed in Scilit:
- Noninvasive Glucose Monitoring in Diabetic Patients: A Preliminary EvaluationClinical Chemistry, 1992
- Multivariate determination of blood substrates in human plasma by FT-NIR spectroscopyPublished by SPIE-Intl Soc Optical Eng ,1992
- Novel design of an ellipsoidal diffuse-reflectance accessory with improved throughputPublished by SPIE-Intl Soc Optical Eng ,1992
- Optical reflectance and transmittance of tissues: principles and applicationsIEEE Journal of Quantum Electronics, 1990
- Calibration modeling by partial least-squares and principal component regression and its optimization using an improved leverage correction for prediction testingChemometrics and Intelligent Laboratory Systems, 1990
- Multivariate determination of glucose in whole blood by attenuated total reflection infrared spectroscopyAnalytical Chemistry, 1989
- Development and application of three-dimensional light distribution model for laser irradiated tissueIEEE Journal of Quantum Electronics, 1987
- Smoothing by spline functionsNumerische Mathematik, 1967
- Die enzymatische Bestimmung von Glucose und Fructose nebeneinanderKlinische Wochenschrift, 1961