Clinical comparison of the ProTon and Tono-Pen tonometers with the Goldmann applanation tonometer.
Open Access
- 1 December 1994
- journal article
- research article
- Published by BMJ in British Journal of Ophthalmology
- Vol. 78 (12) , 895-898
- https://doi.org/10.1136/bjo.78.12.895
Abstract
A clinical evaluation of a new electron ProTon tonometer was performed comparing the values of intraocular pressure (IOP) measured using this instrument with those determined by a similar instrument, Tono-Pen XL, and by Goldmann applanation tonometry. The mean IOP measured in 106 eyes with the ProTon tonometer was not significantly different from that determined with Goldmann applanation, while the IOP values measured with Tono-Pen XL were significantly lower. The 95% limits of agreement between applanation tonometry and ProTon tonometry were between -4 mm Hg and 5 mm Hg and between applanation tonometry and Tono-Pen XL tonometry between -3 mm Hg and 8 mm Hg. The ProTon tonometer appears to have a higher level of accuracy than the Tono-Pen XL tonometer in clinical practice.Keywords
This publication has 16 references indexed in Scilit:
- Tonometry in the general practice setting (I)Acta Ophthalmologica, 2009
- Sources of error with use of Goldmann-type tonometersSurvey of Ophthalmology, 1993
- Evaluation of the Tono-Pen 2 and the X-PertTM Noncontact Tonometers in Cataract SurgeryOphthalmologica, 1993
- Tono-Pen tonometry in normal and in post-keratoplasty eyes.British Journal of Ophthalmology, 1992
- Assessment of Intraocular Pressure in Vitrectomized Gas-Containing EyesArchives of Ophthalmology (1950), 1990
- The Tono-PenArchives of Ophthalmology (1950), 1988
- Comparison of Intraocular Pressure Measurements With the Oculab Tono-Pen vs Manometry in Humans Shortly After DeathAmerican Journal of Ophthalmology, 1988
- Clinical Comparison of the Oculab Tono-Pen to the Goldmann Applanation TonometerOphthalmology, 1987
- Clinical Evaluation of the Oculab Tono-PenAmerican Journal of Ophthalmology, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986