In vivo heating of pacemaker leads during magnetic resonance imaging
Top Cited Papers
Open Access
- 29 November 2004
- journal article
- research article
- Published by Oxford University Press (OUP) in European Heart Journal
- Vol. 26 (4) , 376-383
- https://doi.org/10.1093/eurheartj/ehi009
Abstract
Aims Magnetic resonance imaging (MRI) is well established as an important diagnostic tool in medicine. However, the presence of a cardiac pacemaker is usually regarded as a contraindication for MRI due to safety reasons. In this study, heating effects at the myocardium–pacemaker lead tip interface have been investigated in a chronic animal model during MRI at 1.5 Tesla. Methods and results Pacemaker leads with additional thermocouple wires as temperature sensors were implanted in nine animals. Temperature increases of up to 20°C were measured during MRI of the heart. Significant impedance and minor stimulation threshold changes could be seen. However, pathology and histology could not clearly demonstrate heat-induced damage. Conclusions MRI may produce considerable heating at the lead tip. Changes of pacing parameters due to MRI could be seen in chronic experiments. Potential risk of tissue damage cannot be excluded even though no reproducible alterations at the histological level could be found.Keywords
This publication has 34 references indexed in Scilit:
- Can patients with implantable pacemakers safely undergo magnetic resonance imaging?**Editorials published in the Journal of the American College of Cardiologyreflect the views of the authors and do not necessarily represent the views of JACCor the American College of Cardiology.Journal of the American College of Cardiology, 2004
- Magnetic resonance imaging and cardiac pacemaker safety at 1.5-TeslaPublished by Elsevier ,2004
- Pacing in magnetic resonance imaging environment: Clinical and technical considerations on compatibilityEuropean Heart Journal, 2001
- MR Imaging and Cardiac Pacemakers: In Vitro Evaluation and in Vivo Studies in 51 Patients at 0.5 TRadiology, 2000
- Safe Performance of Magnetic Resonance Imaging on Five Patients with Permanent Cardiac PacemakersPacing and Clinical Electrophysiology, 1996
- Pacemaker MalfunctionAnnals of Internal Medicine, 1993
- Case Report: Nuclear Magnetic Resonance Imaging in a Patient with a PacemakerThe Lancet Healthy Longevity, 1993
- Policies, guidelines, and recommendations for MR imaging safety and patient managementJournal of Magnetic Resonance Imaging, 1992
- Nuclear Magnetic Resonance Imaging in a Patient with a DDD PacemakerPacing and Clinical Electrophysiology, 1989
- The Effects of Nuclear Magnetic Resonance Imagers on External and Implantable Pulse GeneratorsPacing and Clinical Electrophysiology, 1984