Embolic Protection With Filtering or Occlusion Balloons During Saphenous Vein Graft Stenting Retrieves Identical Volumes and Sizes of Particulate Debris

Abstract
Background— Distal embolization of plaque particulate liberated during stenting may cause periprocedural complications. The number, size, and volume of debris released during stenting, however, have not been quantified, rendering embolic protection approaches empiric. We used a novel method of microparticle size assessment to measure volume and characterize individual sizes of particles captured by the PercuSurge GuardWire balloon or a vascular filter during saphenous vein graft stenting. Methods and Results— Braided nitinol filters (average distal pore size 100 μm) were used in 47 saphenous vein grafts in 44 patients. The PercuSurge GuardWire was used in 17 saphenous vein grafts in 16 patients. Particulate debris was subjected to microparticle size analysis (RapidVue, Beckman Coulter). All samples contained particulate debris. For both filter and GuardWire populations, most particles were 3 (range 2 to 84 mm 3 ). Median embolic load per GuardWire was also 16 mm 3 (range 7 to 42 mm 3 ). Histopathologic analysis demonstrated that most samples contained plaque elements and platelet-rich thrombus. Conclusions— During saphenous vein graft interventions, particulate retrieved with a vascular filtering device or an occlusion balloon was similar in amount and character. This supports the notion that unless soluble mediators play an important role in adverse acute clinical events after stenting, the clinical efficacy of filtering devices may be equal to that of occlusion devices.