Randomized Comparison of Distal Protection With a Filter-Based Catheter and a Balloon Occlusion and Aspiration System During Percutaneous Intervention of Diseased Saphenous Vein Aorto-Coronary Bypass Grafts

Abstract
Background— The high rate of periprocedural complications resulting from atherothrombotic embolization after percutaneous intervention in diseased saphenous vein grafts is reduced by distal microcirculatory protection using a balloon occlusion and aspiration system. Whether filter-based catheters, which offer the inherent advantages of maintained perfusion and ease of use, are as effective for this purpose has not been established. Methods and Results— A total of 651 patients undergoing percutaneous intervention of 682 saphenous vein graft lesions were prospectively randomized to distal protection with the filter-based FilterWire EX versus the GuardWire balloon occlusion and aspiration system. Device success was 95.5% and 97.2% with the FilterWire EX and GuardWire, respectively (P=0.25). Postprocedural measures of epicardial flow and angiographic complications were similar between the 2 groups, although bailout IIb/IIIa inhibitors were required slightly less frequently in the FilterWire EX group (0% versu...