Normal lymphocytes from leukemic samples as an internal quality control for fluorescence intensity in immunophenotyping of acute leukemias
Open Access
- 8 November 2005
- journal article
- research article
- Published by Wiley in Cytometry Part B: Clinical Cytometry
- Vol. 70B (1) , 1-9
- https://doi.org/10.1002/cyto.b.20075
Abstract
Background Multiparametric flow cytometry has become an indispensable but complex tool for the diagnosis of acute leukemias. Interpretation of immunophenotypic data within a six-parameter analytical space relies on the standardization and validation of the instrument, the reagents, and the procedure. To address whether or not residual normal lymphocytes, usually present within leukemic samples, can serve as internal quality control for fluorescence intensity, 116 leukemic and 35 normal samples were analyzed. Methods Eight laboratories participated in the study and recruited a total of 151 individuals including 29 patients with B-cell precursor acute lymphoblastic leukemia (BCP-ALL), 77 with acute myeloid leukemia (AML), 10 with T-cell precursor acute lymphoblastic leukemia (T-ALL), and 35 normal bone marrow donors. Lymphocytes were gated according to the CD45hi/SSClo gating strategy, after which median fluorescence intensities (MFI) as well as percentages of positive cells (%positive) for CD19, CD22, CD7, and CD3 were recorded. Nonparametric statistics were used to compare variation within and between laboratories. Results Normal lymphocytes within leukemic samples do not show substantial differences compared to lymphocytes from normal controls with respect to expression of CD19, CD22, CD7, and CD3. In particular, longitudinal control charts of MFI values for CD3 antigen provide useful information on analytical and instrument performance. Conclusion Residual normal lymphocytes can serve as internal quality control for studies addressing fluorescence intensity in the setting of immunophenotyping of acute leukemias.Keywords
This publication has 26 references indexed in Scilit:
- Use of CD45 gating in three and four‐color flow cytometric immunophenotyping: Guideline from the national institute of allergy and infectious diseases, division of AIDSCytometry, 2002
- Frequency difference gating: A multivariate method for identifying subsets that differ between samplesCytometry, 2001
- Flow cytometry in the diagnosis of acute leukemiaSeminars in Hematology, 2001
- SIHONSCORE: a scoring system for external quality control of leukaemia/lymphoma immunophenotyping measuring all analytical phases of laboratory performanceBritish Journal of Haematology, 2001
- Reduction of intra‐ and interlaboratory variation in CD34+ stem cell enumeration using stable test material, standard protocols and targeted trainingBritish Journal of Haematology, 2000
- Quality control of flow cytometric immunophenotyping of haematological malignanciesClinical and Laboratory Haematology, 1999
- Pitfalls in the immunophenotyping of leukaemia and leukaemic lymphomas: survey of 9 years of quality control in The NetherlandsBritish Journal of Haematology, 1996
- Comparison of lymphocyte immunophenotyes obtained simultaneously from two different data acquisition and analysis systems on the same flow cytomete,Cytometry, 1992
- Interlaboratory study of cellular fluorescence intensity measurements with fluorescein‐labeled microbead standardsCytometry, 1991
- A Predictive Value Model for Quality Control: Effects of the Prevalence of Errors on the Performance of Control ProceduresAmerican Journal of Clinical Pathology, 1983