Estimation of Stroke Volume Using Doppler Echocardiography and Left Ventricular Echocardiographic Dimensions in Infants and Children
- 1 October 1986
- journal article
- research article
- Published by Wiley in Acta Paediatrica
- Vol. 75 (s329) , 114-119
- https://doi.org/10.1111/j.1651-2227.1986.tb10397.x
Abstract
To evaluate the accuracy of noninvasive determination of stroke volume in infants and children, 28 patients (age range 4 weeks to 19 years) were studied. Stroke volume was calculated according to Teichholtz from M-mode echocardiographic tracings of left ventricular dimensions in 8 subjects. Agreement with thermodilution performed within 60 min of echocardiography was good (r = 0.995, y = 0.91x + 1.59, SEE = 1.8 ml). Since stroke volume correlated to body size we corrected for (height)3. After this correction there was still good agreement to thermodilution (r = 0.88, y = 1.29x-7.13, SEE = 7.1 ml/H3). M-mode echocardiography was then used as a reference method for evaluating two different Doppler methods in the remaining 20 subjects. Continuous wave Doppler stroke distance, calculated from the mean velocity, was combined with aortic root area (Method I), and stroke distance calculated from maximum velocity was combined with the aortic interleaflet area (Method II). Good agreement was found with Method I (r = 0.95, y = 1.01x-0.14, SEE = 8.1 ml) and Method II (r = 0.95, y = 1.04x-1.14, SEE = 8.4 ml). However, when stroke volume was normalized for (height)3, Method I was found to be superior to Method II.Keywords
This publication has 24 references indexed in Scilit:
- Reproducibility of linear cardiac output measurement by Doppler ultrasound alone.Heart, 1986
- Changes in left ventricular stroke volume measured by Doppler echocardiography.Heart, 1985
- Assessment of cardiac output by the Doppler ultrasound technique alone.Heart, 1985
- Noninvasive Measurement of Cardiac Output by Continuous-Wave Doppler Echocardiography: Initial Experience and Review of the LiteratureMayo Clinic Proceedings, 1984
- Determination of cardiac output by Doppler echocardiography.Heart, 1984
- Determination of cardiac output by transcutaneous continuous-wave ultrasonic Doppler computerThe American Journal of Cardiology, 1984
- Noninvasive Doppler determination of cardiac output in man. Clinical validation.Circulation, 1983
- Transcutaneous Doppler method of measuring cardiac output—IIThe American Journal of Cardiology, 1980
- Transcutaneous aortovelography. A quantitative evaluation.Heart, 1976
- Problems in echocardiographic volume determinations: Echocardiographic-angiographic correlations in the presence or absence of asynergyThe American Journal of Cardiology, 1976