Evaluation of a continuous cardiac output and mixed venous oxygen saturation catheter in critically ill surgical patients
- 1 March 1997
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 25 (3) , 388-391
- https://doi.org/10.1097/00003246-199703000-00003
Abstract
To evaluate the agreement of continuous cardiac output and mixed venous oxygen saturation measurements, obtained with a modified pulmonary artery catheter, with those values obtained by standard intermittent bolus thermodilution and cooximetry. Prospective, clinical investigation. A surgical intensive care unit in a tertiary referral center. Twenty-one adult critically ill surgical patients, requiring pulmonary artery catheter monitoring. None. A modified pulmonary artery catheter capable of continuous monitoring of cardiac output and mixed venous oxygen saturation was used with either an 8.5-Fr or a 9-Fr introducer. At random intervals, the continuous cardiac output measurement was compared with the cardiac output obtained using standard intermittent bolus thermodilution. The system was calibrated every 24 hrs for mixed venous oxygen saturation monitoring. Each saturation obtained by the laboratory cooximeter was compared with that value recorded using the catheter. Data points for 202 pairs of cardiac output (21 patients, 31 catheters) and 65 pairs of mixed venous oxygen saturation (20 patients, 28 catheters) were obtained. The bias and precision of the cardiac output data were 0.49 and 1.01 L/min, respectively. The agreement between the continuous and bolus values decreased as the cardiac output increased. Heart rate did not affect the agreement between the continuous and bolus techniques. The bias and precision of the mixed venous oxygen saturation data were -0.57% and 3.76%, respectively. The hematocrit did not affect the bias or precision of the venous saturation data over the hematocrit range observed (23.2% to 44.6%). Fewer catheter malfunctions were observed when the catheter was used with a 9-Fr introducer than with an 8.5-Fr introducer. The test catheter adequately measures continuous cardiac output and mixed venous oxygen saturation in the clinical setting. Because intermittent bolus thermodilution is not a true "gold standard" for cardiac output determination, new techniques compared with bolus thermodilution may fail to achieve accuracy expectations. A 9-Fr introducer is recommended, as fiberoptic damage may have occurred when the 8.5-Fr introducer was used.Keywords
This publication has 12 references indexed in Scilit:
- Evaluation of a new continuous thermodilution cardiac output monitor in critically ill patientsCritical Care Medicine, 1995
- Is continuous cardiac output measurement using thermodilution reliable in the critically ill patient?Critical Care Medicine, 1994
- A Comparison of Three Pulmonary Artery Oximetry Catheters in Intensive Care Unit PatientsChest, 1992
- Alternatives to Swan-Ganz Cardiac Output MonitoringSurgical Clinics of North America, 1991
- Noninvasive Cardiac Output: Simultaneous Comparison of Two Different Methods with ThermodilutionAnesthesiology, 1990
- Is the Measurement of Cardiac Output Useful in Clinical Practice?British Journal of Anaesthesia, 1988
- Method of assessing the reproducibility of blood flow measurement: factors influencing the performance of thermodilution cardiac output computers.Heart, 1986
- Major Errors in Thermodilution Cardiac Output Measurement during Rapid Volume InfusionAnesthesiology, 1985
- Doppler echocardiographic measurement of cardiac output: Comparison with a non-golden standardThe American Journal of Cardiology, 1984
- Thermodilution cardiac output: A critical analysis and review of the literatureJournal of Surgical Research, 1979