Analysis of the detective quantum efficiency of a developmental detector for digital mammography
- 11 November 1999
- journal article
- Published by Wiley in Medical Physics
- Vol. 26 (11) , 2273-2285
- https://doi.org/10.1118/1.598741
Abstract
We are developing a modular detector for applications in full field digital mammography and for diagnostic breast imaging. The detector is based on a design that has been refined over the past decade for applications in x-ray crystallography [Kalata et al., Proc. SPIE 1345, 270-279 (1990); Phillips et al. ibid. 2009, 133-138 (1993), Phillips et al., Nucl. Instrum. Methods Phys. Rev. A 334, 621-630 (1993)]. The full field mammographic detector, currently undergoing clinical evaluation, is formed from a 19 cm x 28 cm phosphor screen, read out by a 2 x 3 array of butted charge-coupled device (CCD) modules. Each 2k x 2k CCD is optically coupled to the phosphor via a fiber optic taper with dimensions of 9.4 cm x 9.4cm at the phosphor. This paper describes the imaging performance of a two-module prototype, built using a similar design. In this paper we use cascaded linear systems analysis to develop a model for calculating the spatial frequency dependent noise power spectrum (NPS) and detective quantum efficiency (DQE) of the detector using the measured modulation transfer function (MTF). We compare results of the calculation with the measured NPS and DQE of the prototype. Calculated and measured DQEs are compared over a range of clinically relevant x-ray exposures and kVps. We find that for x-ray photon energies between 10 and 28 keV, the detector gain ranges between 2.5 and 3.7 CCD electrons per incident x-ray, or approximately 5-8 electrons per absorbed x ray. Using a Mo/Mo beam and acrylic phantom, over a detector entrance exposure range of approximately 10 to 80 mR, the volume under the measured 2-d NPS of the prototype detector is proportional to the x-ray exposure, indicating quantum limited performance. Substantial agreement between the calculated and measured values was obtained for the frequency and exposure dependent NPS and DQE over a range of tube voltage from 25 to 30 kVp.Keywords
This publication has 31 references indexed in Scilit:
- Digital radiology using active matrix readout of amorphous selenium: Construction and evaluation of a prototype real‐time detectorMedical Physics, 1997
- Digital radiology using active matrix readout of amorphous selenium: Theoretical analysis of detective quantum efficiencyMedical Physics, 1997
- Study of afterglow in x-ray phosphors for use on fast-framing charge-coupled device detectorsOptical Engineering, 1997
- Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat‐panel imagers (AMFPIs) for diagnostic radiologyMedical Physics, 1997
- A spatial‐frequency dependent quantum accounting diagram and detective quantum efficiency model of signal and noise propagation in cascaded imaging systemsMedical Physics, 1994
- The detective quantum efficiency of CCD and vidicon-based detectors for X-ray crystallographic applicationsJournal of Applied Crystallography, 1992
- A simple method for determining the modulation transfer function in digital radiographyIEEE Transactions on Medical Imaging, 1992
- Charge-Coupled-Device Charge-Collection Efficiency And The Photon-Transfer TechniqueOptical Engineering, 1987
- Detective quantum efficiency of imaging systems with amplifying and scattering mechanismsJournal of the Optical Society of America A, 1987
- Image fluctuations in cascade intensifiersBritish Journal of Applied Physics, 1959