Cross-calibration, precision and patient dose measurements in preparation for clinical trials using dual energy X-ray absorptiometry of the lumbar spine.
- 1 April 1999
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 72 (856) , 354-362
- https://doi.org/10.1259/bjr.72.856.10474496
Abstract
Large, multicentre clinical trials using DXA to monitor bone density following intervention are now common. At the same time, several different bone densitometers and calibration phantoms are currently in use. The aim of this study was to document the technical information required on cross-calibration of equipment, reproducibility and patient dose before commencing a multicentre clinical trial. To this end, we obtained an in vitro and in vivo cross-calibration of two machines (a Hologic QDR 2000 and a Lunar DPX-L) that were not significantly different. Interobserver and intraobserver precision, and radiation dose were also measured and three commonly used phantoms assessed for their usefulness in cross-calibration and quality assurance. Measured in vitro precision on the two machines (0.3-0.7%) was better than that specified by the manufacturers. In vivo precision was worse (1.4-2.1%), as might be expected in patients with reduced bone mass. Mean entrance skin radiation doses on each machine were 280 microSv for the QDR 2000 and 38 microSv for the DPX-L. No one phantom is ideal, but the European Spine Phantom or Lunar Aluminium Spine Phantom will provide an adequate cross-calibration for a clinical trial. This study demonstrates that an adequate cross-calibration can be obtained for use in groups of patients and that the equipment used is reproducible with a low radiation output.Keywords
This publication has 12 references indexed in Scilit:
- Long-term quality control of DXA: A comparison of shewhart rules and cusum chartsOsteoporosis International, 1997
- Patient dose in dual x-ray absorptiometryOsteoporosis International, 1994
- Ultrasonic velocity measurements through the calcaneus: Which velocity should be measured?Osteoporosis International, 1993
- Enhanced precision with dual-energy x-ray absorptiometryCalcified Tissue International, 1992
- Comparative performance in vitro and in vivo of Lunar DPX and Hologic QDR-1000 dual energy X-ray absorptiometersThe British Journal of Radiology, 1991
- Guidelines for clinical trials in OsteoporosisOsteoporosis International, 1991
- In vivo and in vitro precision for bone density measured by dual-energy X-ray absorptionOsteoporosis International, 1991
- Dual-energy radiographic absorptiometry of the lumbar spine: clinical experience with two different systems.Radiology, 1990
- Effective dose equivalent in dual X-ray absorptiometryThe British Journal of Radiology, 1990