Low wear and friction in alumina/alumina total hip joints: A hip simulator study
Open Access
- 1 January 1998
- journal article
- research article
- Published by Medical Journals Sweden AB in Acta Orthopaedica
- Vol. 69 (5) , 443-448
- https://doi.org/10.3109/17453679808997776
Abstract
We studied the wear of state-of-the-art alumina/alumina total hip joints with the uniaxial five-station, and the three-axis single-station hip joint simulators of Helsinki University of Technology, The diameters of the joints were 26, 28 and 32 mm. Visual examination and weighing of the components showed no wear. Scanning electron microscopy showed that (a) the wear marks were slight, consisting of removal of grains, multidirectional grooving and mild abrasion, (b) the wear mechanisms were similar to those seen in components removed from patients, and (c) there was no marked difference in the specimens worn in the uniaxial vs. three-axis simulator. The coefficient of friction was measured throughout the 5 million cycle wear test with the three-axis simulator. The average value was 0.007. We conclude that alumina/alumina joints had low wear and friction under the present test conditions. The results are promising, since minimal wear is one of the prerequisites for long-term success in prosthetic joints.Keywords
This publication has 7 references indexed in Scilit:
- Start up and steady state friction of alumina against aluminaWear, 1997
- A Three-Axis Hip Joint Simulator for Wear and Friction Studies on Total Hip ProsthesesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1996
- The safety of ceramic balls on metal stems in hip arthroplastyAdvanced Materials, 1994
- Examination of Alumina Ceramic Components from Total Hip ArthroplastiesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1993
- A Five-Station Hip Joint Simulator for Wear Rate StudiesProceedings of the Institution of Mechanical Engineers, Part H: Journal of Engineering in Medicine, 1992
- On the Material and the Tribology of Alumina-Alumina Couplings for Hip Joint ProsthesesPublished by Wolters Kluwer Health ,1992
- Role of Ceramic ImplantsClinical Orthopaedics and Related Research, 1992