Surface Roughness of Ceramic Femoral Heads After in Vivo Transfer of Metal
- 1 March 2005
- journal article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 87 (3) , 577-582
- https://doi.org/10.2106/jbjs.d.01790
Abstract
Background: A dark metallic-appearing smear, resembling a lead pencil mark, may be seen on a ceramic femoral head component at revision total hip arthroplasty. The purpose of this study was to investigate the hypothesis that such a mark on a retrieved ceramic femoral head is associated with increased surface roughness of the head and increased polyethylene liner wear in total hip replacement. Methods: Fifteen ceramic prosthetic femoral heads retrieved from fifteen patients at revision arthroplasty were examined in this study. Thirteen heads had been in vivo for an average of 10.8 years (range, 7.8 to 14.2 years). The remaining two heads had been in vivo for less than one month. The surface roughness characteristics of the explanted ceramic heads, the linear wear of the polyethylene liner, and the patient activity levels after the primary replacement and before the revision were determined. Results: Four of the thirteen ceramic heads that had been in vivo for ≥7.8 years had severe smears (>6% of the surface area), and the remaining nine heads had slight smears (Conclusions: The results of this study confirm the hypothesis that a visual dark metallic-appearing smear on a ceramic femoral head correlates with increased surface roughness of the head and increased polyethylene wear. These findings imply that contact of a ceramic femoral head with a metallic material, such as may occur with femoral head reduction or dislocation of a total hip replacement, is best avoided to prevent this metallic smear phenomenon.Keywords
This publication has 17 references indexed in Scilit:
- A comparison of polyethylene wear in hips with cobalt-chrome or zirconia headsThe Journal of Bone and Joint Surgery. British volume, 2001
- Poor outcome of 44 cemented total hip arthroplasties with alumina ceramic heads: Clinical evaluation and retrieval analysis after 10-16 yearsActa Orthopaedica, 2001
- Modern Ceramic-on-Ceramic Total Hip Systems in the United StatesClinical Orthopaedics and Related Research, 2000
- Comparison of in vivo wear between polyethylene liners articulating with ceramic and cobalt-chrome femoral headsThe Journal of Bone and Joint Surgery. British volume, 2000
- Ceramic Bearing SurfacesPublished by Wolters Kluwer Health ,1999
- Biologic anchorage of cruciate ligament prosthesis: Bone ingrowth and fixation of the Gore-Tex® ligament in sheepActa Orthopaedica, 1993
- Wear of the polyethylene acetabular cup: Metallic and ceramic heads compared in a hip simulatorActa Orthopaedica, 1993
- Ten-year socket wear in 66 hip arthroplasties: Ceramic versus metal headsActa Orthopaedica, 1990
- The effect of the surface finish of alumina and stainless steel on the wear rate of UHMW polyethyleneBiomaterials, 1986