Dosimetric IMRT verification with a flat‐panel EPID
- 19 November 2003
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 30 (12) , 3143-3155
- https://doi.org/10.1118/1.1625440
Abstract
A convolution-based calibration procedure has been developed to use an amorphous silicon flat-panel electronic portal imaging device (EPID) for accurate dosimetric verification of intensity-modulated radiotherapy (IMRT) treatments. Raw EPID images were deconvolved to accurate, high-resolution 2-D distributions of primary fluence using a scatter kernel composed of two elements: a Monte Carlo generated kernel describing dose deposition in the EPID phosphor, and an empirically derived kernel describing optical photon spreading. Relative fluence profiles measured with the EPID are in very good agreement with those measured with a diamond detector, and exhibit excellent spatial resolution required for IMRT verification. For dosimetric verification, the EPID-measured primary fluences are convolved with a Monte Carlo kernel describing dose deposition in a solid water phantom, and cross-calibrated with ion chamber measurements. Dose distributions measured using the EPID agree to within 2.1% with those measured with film for open fields of 2 x 2 cm2 and 10 x 10 cm2. Predictions of the EPID phantom scattering factors (SPE) based on our scatter kernels are within 1% of the SPE measured for open field sizes of up to 16 x 16 cm2. Pretreatment verifications of step-and-shoot IMRT treatments using the EPID are in good agreement with those performed with film, with a mean percent difference of 0.2 +/- 1.0% for three IMRT treatments (24 fields).Keywords
Funding Information
- Alberta Cancer Board (R-484)
This publication has 31 references indexed in Scilit:
- Fast and accurate leaf verification for dynamic multileaf collimation using an electronic portal imaging deviceMedical Physics, 2002
- Dosimetric verification of inverse planned step and shoot multileaf collimator fields from a commercial treatment planning systemJournal of Applied Clinical Medical Physics, 2002
- Monte Carlo calculation of nine megavoltage photon beam spectra using the BEAM codeMedical Physics, 2002
- An iterative EPID calibration procedure for dosimetric verification that considers the EPID scattering factorMedical Physics, 2001
- Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging deviceMedical Physics, 2001
- A two-step algorithm for predicting portal dose images in arbitrary detectorsMedical Physics, 2000
- Dosimetric verification of intensity‐modulated fieldsMedical Physics, 1996
- Transmission dosimetry with a liquid-filled electronic portal imaging deviceInternational Journal of Radiation Oncology*Biology*Physics, 1996
- Portal dose measurement in radiotherapy using an electronic portal imaging device (EPID)Physics in Medicine & Biology, 1995
- Portal imaging technology: Past, present, and futureSeminars in Radiation Oncology, 1995